Acceso abierto

Application of recombination chambers for determination of neutron ambient dose equivalent at the door to the maze of linear accelerator vault


Cite

Recombination chambers and 2202D Studsvik-Alnor reference remmeter were used for determination of ambient dose equivalent, H*(10) and its neutron and gamma components at the door to the maze of 15 MV medical accelerator. The measurements were performed at opened and partly closed door. The results showed that the ambient dose equivalent rate was 76.2 μSv h-1 at the opened door and only 2.7 μSv h-1 at slightly opened door, 2.5 m from the entrance to the maze.