Does CyberKnife improve dose distribution versus IMRT and VMAT on a linear accelerator in low-risk prostate cancer?
Categoría del artículo: Research Article
Publicado en línea: 28 mar 2022
Páginas: 259 - 266
Recibido: 21 oct 2021
Aceptado: 31 ene 2022
DOI: https://doi.org/10.2478/raon-2022-0010
Palabras clave
© 2022 Dorota Maria Borowicz, Agnieszka Skrobała, Marta Kruszyna-Mochalska, Julian Malicki, published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Background
Hypofractionated stereotactic body radiation therapy (SBRT) for prostate cancer (PCa) can be delivered with the robot-assisted CyberKnife (CK) system or on a linear accelerator using dynamic intensity-modulated radiotherapy (IMRT) or volumetric arc radiotherapy (VMAT). This retrospective study was performed to determine whether CK offers better dose distribution than IMRT and/or VMAT.
Materials and methods
Treatment plans for three techniques were prepared using the same treatment parameters (36.35 Gy, 7.25 Gy/fr). We evaluated target coverage, conformity index (CI), homogeneity index (HI), gamma index (GI), and organs at risk (OAR) constraints.
Results
The mean planning target volume (PTV) dose for CK (39.58 Gy) was significantly greater than VMAT or IMRT (both 36.25 Gy). However, CK resulted in a wider dose range (31.48 to 45.89 Gy)
Conclusions
For most variables, CK was superior to both VMAT and IMRT. However, dynamic IMRT techniques, especially VMAT, do not differ significantly from CK plans and are therefore acceptable alternatives to CyberKnife.