Biomechanical Uniaxial Analysis of Porcine Tendon in the CellScale BioTester® 5000
Publicado en línea: 21 sept 2023
Páginas: 23 - 27
Recibido: 20 jun 2023
Aceptado: 27 jun 2023
DOI: https://doi.org/10.2478/jim-2023-0009
Palabras clave
© 2023 Alexandru Fofiu et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
Background
The study was aimed to evaluate whether a mechanical biaxial tester can be used in a uniaxial mode to evaluate the mechanical properties of tendons.
Materials and methods
The study was carried out on specimens of porcine superficial digital flexor tendon (n = 9). The mechanical properties (elastic modulus, and stress at 15% strain) were measured two times consecutively in the uniaxial mode with the BioTester® 5000 (CellScale) equipment.
Results
Values of 0.313 ± 0.096 MPa for the elastic (Young’s) modulus and of 0.702 ± 0.174 MPa for the stress (at 15% strain) were measured, indicating that the porcine superficial digital flexor tendon is not a strong tendon.
Conclusions
When suitable specimens cannot be obtained for a biaxial evaluation, tendons can be evaluated mechanically in the BioTester® 5000 employing the uniaxial mode.