Group | Compound | Chemical structure | IARC |
FDA |
WHO study group |
F |
||
---|---|---|---|---|---|---|---|---|
TACI | TNCRI | CRI | ||||||
Alkaloids | Nicotine CAS 54-11-5 |
|
– | – | – | – | – | |
Aldehydes | Acetaldehyde CAS 75-07-0 |
|
Group 2B, 1999 | CA, RT, AD | 6.1 | 67.1 | 9.18×10−5 | |
Acrolein CAS 107-02-8 |
|
Group 2A, 2021 | RT, CT | – | 1099 | – | ||
Formaldehyde CAS 50-000-0 |
|
Group 1, 2012 | CA, RT | – | 19.8 | 9.90×10−6 | ||
Crotonaldehyde CAS 4170-30-3 |
|
Group 2B, 2021 | CA | – | – | – | ||
Propionaldehyde CAS 123-38-6 |
|
– | CT, RT | – | – | – | ||
|
– | – | – | – | – | |||
Aromatic amines | 3-Aminobiphenyl CAS 2243-47-2 |
|
– | – | – | – | – | |
4-Aminobiphenyl CAS 92-67-1 |
|
Group 1, 2012 | CA | – | – | 3.60×10−7 | ||
1-Aminonaphthalene (1-Naphthylamine) CAS 134-32-7 |
|
Group 3, 1987 | CA | – | – | – | ||
2-Aminonaphthalene (2-Naphthylamine) CAS 91-59-8 |
|
Group 1, 2012 | CA | – | – | 1.80×10−7 | ||
Hydrocarbons | Benzene CAS 71-43-2 |
|
Group 1, 2018 | CA, CT, RTD | 2.6 | 0.64 | 6.71×10−5 | |
1,3–Butadiene CAS 106-99-0 |
|
Group 1, 2012 | CA, CT, RTD | 9.9 | 2.4 | 3.02×10−4 | ||
Isoprene CAS 78-79-5 |
|
Group 2B, 1999 | CA | – | – | – | ||
Styrene CAS 100-42-5 |
|
Group 2A, 2019 | CA, RT | – | 0.01 | – | ||
Toluene CAS 108-88-3 |
|
Group 3, 1999 | RT, RTD | – | 0.22 | – | ||
Polycyclic aromatic hydrocarbons | Benzo[ |
|
Group 1, 2012 | CA | 0.0086 | – | 1.93×10−6 | |
Tobacco-specific nitrosamines | 4-( |
|
Group 1, 2012 | CA | 3.4 | – | 7.80×10−6 | |
|
Group 1, 2012 | CA | 0.29 | – | 3.80×10−5 | |||
|
Group 3, 2007 | – | – | – | ||||
|
Group 3, 2007 | – | – | – | ||||
Phenols | Catechol CAS 120-80-9 |
|
Group 2B, 1999 | CA | 0.58 | 1.2 | – | |
|
– | CA, RT | – | 0.01 | – | |||
|
– | CA, RT | – | 0.01 | – | |||
|
– | CA, RT | – | 0.01 | – | |||
Phenol CAS 108-95-2 |
|
Group 3, 1999 | CA, RT | – | 0.07 | – | ||
Hydroquinone CAS 123-31-9 |
|
Group 3, 1999 | – | 1.2 | – | – | ||
Resorcinol CAS 108-46-3 |
|
Group 3, 1999 | – | – | – | – | ||
Other organic compounds | Acetone CAS 67-64-1 |
|
– | RT | – | – | – | |
Acrylonitrile CAS 107-13-1 |
|
Group 2B, 1999 | CA, RT | 1.4 | 2.1 | 1.29×10−4 | ||
Quinoline CAS 91-22-5 |
|
Group 2B, 1999 | CA | – | – | – | ||
Pyridine CAS 110-86-1 |
|
Group 2B, 2019 | – | – | – | – | ||
Metal and metalloids | Arsenic CAS 740-38-2 | As | Group 1, 2012 | CA, CT, RT | – | 0.16 | 1.16×10−4 | |
Cadmium CAS 7440-34-9 | Cd | Group 1, 2012 | CA, RT, RDT | 1.7 | 2.6 | 2.16×10−5 | ||
Chromium CAS 7440-47-3 | Cr | Group 3, 1990 | CA, RT, RDT | – | – | 3.15×10−5 | ||
Lead CAS 7439-92-1 | Pb | Group 2B, 1987 | CA, CT, RDT | 0.00 | – | 7.68×10−9 | ||
Mercury CAS 7439-97-6 | Hg | Group 3, 1993 | CA, RDT | – | 0.02 | – | ||
Nickel CAS 7440-02-0 | Ni | Group 2B, 1990 | CA, RT | – | – | 1.43×10−7 | ||
Selenium CAS 7782-49-2 | Se | Group 3, 1987 | RT | – | – | – | ||
Other constituents | Ammonia CAS 7664-41-7 | NH3 | – | RT | – | 0.07 | – | |
Carbon monoxide CAS 630-08-0 | CO | – | RDT | – | 1.3 | – | ||
Hydrogen cyanide CAS 74-90-8 | HCN | – | RT, CT | – | 17.2 | – | ||
Other constituents | Nitrogen oxides CAS 10102-43-9 | NOn | – | – | – | 3.1 | – |
SOP 08. Standard Operating Procedure for Determination of Aldehydes in Mainstream Cigarette Smoke Under ISO and Intense Smoking Conditions. 31 August 2018 ( |
CRM No. 74. Determination of Selected Carbonyls in Mainstream Cigarette Smoke by HPLC (August 2019) ( |
|||||
---|---|---|---|---|---|---|
Formaldehyde, acetaldehyde, acrolein (acrylaldehyde) | Formaldehyde, acetaldehyde, acetone, acrolein, propionaldehyde, crotonaldehyde, 2-butanone and |
|||||
Cigarettes are smoked on a smoking machine according to the ISO or intense regime. One cigarette is smoked per smoke trap, containing 300 mg CX-572 particles and a glass fibre filter pad. This standard operating procedure is suitable for the quantitative determination of the following three aldehydes in mainstream (MS) cigarette smoke: formaldehyde, acetaldehyde, and acrolein to be analyzed by liquid chromatography. The aldehydes in mainstream tobacco smoke are extracted by adding a solution containing a mixture of carbon disulphide and methanol and are derivatized with a solution of 2,4-dinitrophenylhydrazine (DNPH). The samples are subjected to analysis using high-performance liquid chromatography (HPLC), equipped with ultraviolet (UV) or diode array detector (DAD). | Cigarettes are smoked on a smoking machine that has been fitted with impingers, but without the specified filter pad holder containing the glass fiber filter. The carbonyls in mainstream tobacco smoke are trapped by passing each puff through an impinger device containing an acidified solution of 2,4-dinitrophenylhydrazine (DNPH) in acetonitrile. An aliquot of the smoke extract is then syringe-filtered and diluted with 1% Trizma™ base in aqueous acetonitrile. The samples are subjected to analysis using reverse phase HPLC/UV or HPLC/DAD. | |||||
Linear smoking machine | Linear smoking machine | |||||
Rotary smoking machine | Rotary smoking machine | |||||
HPLC system with a UV and/or DAD detector | HPLC system with a UV and/or DAD detector | |||||
Carboxen 572 particles, 20/45 mesh size | ||||||
|
|
|||||
DNPH solution: | DNPH solution: | |||||
Approximately 1 g DNPH-HCl is added to a 50-mL volumetric flask, then 10 mL of 85% phosphoric acid is carefully added. | Approximately 150 mL deionized water is added to a 200-mL volumetric flask, then 28 mL of 85% phosphoric acid is carefully added. The solution is made up to a volume of 200 mL with deionized water. 6.8 g (0.024 mole) of DNPH (approximately 30% water) is weighed into a 2-L amber volumetric flask and 1 L of acetonitrile is added. 58 mL of the diluted phosphoric acid solution is added and gently mixed. The volume is diluted with deionized water. | |||||
Trizma™ base dilution solution: | ||||||
MeCN : 1% aqueous Trizma™ (80:20 |
||||||
The calibration should cover the concentration range of interest. The linear range is approximate to the aldehyde content in tobacco smoke. | The calibration should cover the concentration range of interest. The linear range is approximate to the carbonyl content in tobacco smoke. | |||||
ISO / Intense smoking regime | ISO / Intense smoking regime | |||||
Linear smoking machine: 1 | Linear smoking machine: 2 | |||||
Rotary smoking machine: 1 | Rotary smoking machine: 5 | |||||
The aldehydes mainstream apparatus is assembled on the smoking machine using Carboxen 572 and glass fibre filter pad. The method is used for simultaneous determination of aldehydes (SOP 08) and volatile organics (SOP 09) in tobacco smoke. Transfer 0.5 mL to 5 mL flask and add 200 μL of DNPH solution and diluted to 5 mL. An aliquot of solution is analyzed by HPLC/UV or HPLC/DAD. | The carbonyl mainstream apparatus is assembled on the smoking machine without using the filter pads and filter holders and cigarettes are smoked. The DNPH smoke extract solution is allowed to sit for five to thirty minutes. After that 6 mL of 1% Trizma™ base solution is pipetted into a 10 mL volumetric flask. An aliquot of solution is analyzed by reverse-phase HPLC/UV or HPLC/DAD. | |||||
High performance liquid chromatography with UV or DAD Detector | High performance liquid chromatography with UV or DAD Detector (RP HPLC - UV or RP HPLC - DAD). | |||||
LC column: | Ascentis RP-Amide (150 mm × 4.6 mm, 3 μm) or equivalent | LC column: | 250 mm x 4 mm, Reversed Phase (RP) C18 (5 μm), or equivalent | |||
Column Temperature: | 30 ± 5 °C | Disposable Guard Column: | 4 mm x 4 mm, RP C18 (5 μm), or equivalent | |||
Injection volume: | 20 μL | |||||
UV detection: | 360 nm or maximum wavelength at 300–400 nm | Column Temperature: | 30 °C | |||
Injection volume: | 20 μL | |||||
Degasser: | On | UV or DAD detection: | 365 nm | |||
Binary pomp flow: | 1.0 mL/min | Flow rate: | 1.5 mL/min | |||
Mobile Phase A: | water | Mobile Phase: | ||||
Mobile Phase B: | acetonitrile |
Solvent A: Prepare 2 L of 30% Acetonitrile, 10% THF, 1% IPA in Type I water, filter and degas (UHP Helium sparged) Solvent B: Prepare 2 L of 65% Acetonitrile, 1% THF, 1% IPA in Type I water, filter and degas (UHP Helium sparged) Solvent C: Acetonitrile (UHP Helium sparged) |
||||
Total analysis time: | 45 min | |||||
Gradient conditions: | ||||||
0.0 | 100% A | 0% B | ||||
10.0 | 55% A | 45% B | ||||
25.0 | 55% A | 45% B | Gradient conditions: | |||
30.0 | 0% A | 100% B | 0.0 | 100% A | 0% B | 0% C |
30.0 | 0% A | 100% B | 8.0 | 70% A | 30% B | 0% C |
31.0 | 55% A | 45% B | 20.0 | 47% A | 53% B | 0% C |
40.0 | 55% A | 45% B | 27.0 | 0% A | 100% B | 0% C |
30.0 | 0% A | 0% B | 100% C | |||
32.0 | 0% A | 0% B | 100% C | |||
34.0 | 95% A | 5% B | 0% C | |||
Equilibration 10.0 | 100% A | 0% B | 0% C |
Method description | |||||
---|---|---|---|---|---|
Hydroquinone, resorcinol, catechol, phenol, |
|||||
Cigarettes are smoked on a routine smoking machine and mainstream smoke is trapped on a glass-fibre filter pad (GFP). The GFP is extracted with aqueous acetic acid. An aliquot of the extract is syringe filtered and analysed by HPLC/FLD. | |||||
Linear smoking machine or Rotary smoking machine HPLC/FLD | |||||
1% acetic acid solution | |||||
The calibration should cover the concentration range of interest. The linear range is approximate to the phenolic compounds in tobacco smoke. | |||||
ISO smoking regime | |||||
Linear smoking machine: 5 | |||||
Rotary smoking machine: 10 | |||||
Extraction of filter pads: | |||||
The pads from the holders are removed, transferred into an Erlenmeyer flask and 40 mL 1% acetic acid solution (5 cigarettes) and 80 mL 1% acetic acid solution (10 cigarettes) is added. The extraction of phenolic compounds from the pad is until the CFP has disintegrated. The CFP extracts may need to be diluted. | |||||
HPLC/FLD | |||||
HPLC column with pentafluoro phenylpropyl (PFP) stationary phase; an example of the column dimensions are 3-μm, a 150-mm column and 4.6 mm internal diameter or equivalent. | Wavelength programmable fluorescence detector settings (example): | ||||
Time (min) | Excitation (nm) | Emission (nm) | |||
Column Temperature: | Ambient | 0.0 | 280 | 310 | |
Auto-sampler Tray Temperature: | 4 °C | 12.4 | 280 | 310 | |
Injection volume: | 10 μL or 20 μL | 12.5 | 274 | 298 | |
23.0 | 274 | 298 | |||
Mobile Phases: | 24.0 | 280 | 310 | ||
Mobile phase A: | Prepare 2 L of 1% acetic acid in deionized water and degas | 28.0 | 280 | 310 | |
Mobile phase B: | Prepare 2 L of 1% acetic acid in methanol and degas | ||||
Flow rate: | 0.8 mL/min | ||||
The following is an example of gradient conditions: | |||||
Time (min) | % A | % B | |||
0 | 78 | 22 | |||
8 | 78 | 22 | |||
8.5 | 55 | 45 | |||
21 | 55 | 45 | |||
22 | 0 | 100 | |||
28 | 0 | 100 |
Method description | ||||
---|---|---|---|---|
1-Aminonaphthalene, 2,6-dimethylaniline, 2-aminonaphthalene, 3-aminobiphenyl, 4-aminobiphenyl, |
||||
2,6-Dimethylaniline- |
||||
Cigarettes are smoked on a standard smoking machine. The mainstream smoke is collected on a glass-fibre filter pad (GFP). After addition of internal standards, the GFP is extracted with dichloromethane using a laboratory shaker for 20 min. The extract is filtered, derivatized with HFBA, purified on a Florisil SPE and analysed by GC/MS-NCI. | ||||
Linear smoking machine or Rotary smoking machine GC/MS-NCI | ||||
Dichloromethane | ||||
1-Aminonaphthalene: | 0.4 – 5.0 ng/mL | |||
3-Aminobiphenyl: | 0.07 – 0.9 ng/mL | |||
2,6-Dimethylaniline: | 0.1 – 3.1 ng/mL | |||
4-Aminobiphenyl: | 0.07 – 0.9 ng/mL | |||
2-Aminonaphthalene | 0.4 – 5.0 ng/mL | |||
0.1 – 1.1 ng/mL | ||||
1 – 16 ng/mL | ||||
ISO smoking regime | Intense smoking regime | |||
• Linear smoking machine: 5 | • Linear smoking machine: 2–3 | |||
• Rotary smoking machine: 10 | • Rotary smoking machine: 5 | |||
Extraction of filter pads: | Sample derivatization: | Sample clean-up: | ||
The pads are removed from the holders, transferred into the Erlenmeyer flask and 50 mL dichloromethane + 1 mL internal standard intermediate solution is added. The extraction of aromatic amines from the pad is for 40 min. | 5 mL aliquot is derivatized with HFBA (25 μL for ISO regime and 50 μL for intense regime) for 40 min. | SPS cartridge (Florisil) is pre-conditioned with 12 mL of dichloromethane (discard). Whole sample extract is transferred into the cartridge and is allowed to pass through the cartridge (collect). The cartridge is eluted with 8.5 mL dichloromethane (collect). The both eluents are combined, mixed well and an aliquot is transferred to a GC vial for GC/MS-NCI analysis. | ||
GC equipment and its operating conditions: | Mass detection conditions: | |||
GC column: | low/mid-polarity, with a (14%-cyanopropylphenyl)-methylpolysiloxane stationary phase, a 30-m column with a 0.32-mm internal diameter and 1-μm film thickness or equivalent. | Interface/transfer line temperature: | 240 °C (ion trap 200 °C) | |
MS source temperature: | 150 °C | |||
MS quadrupole temperature: | 106 °C | |||
MS mode: | NCI | |||
Data acquisition mode: | Selected Ion Monitoring (SIM) | |||
Injector temperature: | 250 °C | Reagent gas: | Methane at 40% flow | |
Injection mode: | Splitless | Ion peak for identification (m/z): | ||
Injection volume: | 3 μL | 283 | ||
Mode: | Constant flow | 2,6-Dimethylaniline: | 297 | |
Flow rate: | 1.5 mL/min | 299 | ||
Column temperature: | 40 °C (0.5 min), 15 °C/min to 240 °C (hold for 5 min), 50 °C/min to 270 °C (hold for 10 min) | 1-Aminonaphthalene: | 319 | |
2-Aminonaphthalene: | 319 | |||
3-Aminobiphenyl: | 345 | |||
4-Aminobiphenyl: | 345 | |||
290 | ||||
2,6-Dimethylaniline- |
306 | |||
2-Aminonaphthalene- |
326 | |||
4-Aminobiphenyl- |
354 |
SOP 05. Standard Operating Procedure for Determination of Benzo[ |
CRM No. 58. Determination of Benzo[ |
||||
---|---|---|---|---|---|
Benzo[ |
Benzo[ |
||||
B[ |
B[ |
||||
Mainstream smoke total particulate matter from the cigarette sample is trapped onto a GFP. Internal standard is spiked onto the GFP, which is extracted with cyclohexane. The cyclohexane extract is purified using SPE by passing through a pure silica unbound phase SPE cartridge, and the eluent is collected. The sample is analyzed by GC/MS with electron ionization detection. | Mainstream smoke total particulate matter from the cigarette sample is trapped onto a GFP. Internal standard is spiked onto the GFP, which is extracted with methanol, and the methanol extract is diluted with water. Crude water/methanol smoke extract is purified using SPE by passing through a cyclohexyl bonded silica (CH) SPE cartridge, followed by the elution of B[ |
||||
Linear smoking machine or Rotary smoking machine GC/MS | Linear smoking machine or Rotary smoking machine GC/MS | ||||
Cyclohexane | Methanol | ||||
2 – 60 ng/mL | 1 – 200 ng/mL | ||||
ISO smoking regime | ISO smoking regime | ||||
• Linear smoking machine: 5 | • Linear smoking machine: 5 – 10 | ||||
• Rotary smoking machine: 20 | • Rotary smoking machine: 10 – 20 | ||||
Intense smoking regime | Intense smoking regime | ||||
• Linear smoking machine: 3 | • Linear smoking machine: 5 – 10 | ||||
• Rotary smoking machine: 10 | • Rotary smoking machine: 10 – 20 | ||||
Extraction of filter pads: The pads from the holders are removed, transferred into Erlenmeyer flask and cyclohexane and internal standard (40 mL cyclohexane + 40 μL of the B[ |
Extraction of filter pads: The pads from the holders are removed, transferred into Erlenmeyer flask and methanol and internal standard (20 mL + 200 μL of the B[ |
||||
Sample clean-up: SPE cartridge is pre-conditioned with 10 mL cyclohexane (discard). 10 mL of sample extract is pipetted onto the cartridge and is allowed to pass through the cartridge (collect). The cartridge is eluted with two further 15-mL aliquots of cyclohexane, allowing the cartridge to run dry after the last aliquot is passed through. The cyclohexane solution is evaporated almost to dryness and is dissolved with 1 mL cyclohexane. | Sample clean-up: SPE cartridge is pre-conditioned with 10 mL methanol and 10 mL of a mixture of water and methanol (60:40, |
||||
GC/MS | GC/MS | ||||
GC Column - Fused silica capillary column with a methylphenyl (5%) polysiloxane stationary phase, a 30-m column with a 0.25-mm internal diameter and 0.25-μm film thickness is suitable | GC Column - Fused silica capillary column with a methylphenyl (5%) polysiloxane stationary phase, 30-m column with a 0.25 mm internal diameter and 0.25 μm film thickness is suitable. | ||||
Injector temperature: | 280 °C (ion trap: 200 °C) | Injector temperature: | 290 °C | ||
Injection mode: | Splitless | Injection mode: | Splitless | ||
Injection volume : | 1 μL or 2 μL | Injection volume: | 1 μL | ||
Mode: | Constant flow | Mode: | Constant flow | ||
Flow rate: | 1.2 mL/min (ion trap 1.0 mL/min) | Flow rate: | 0.9 mL/min | ||
Column temperature: | 150 °C for 0 min; 6 °C/min to 260 °C, hold at 260 °C for 7 min 50 °C/min to 290 °C, hold at 290 °C for 20 min | Column temperature: | 80 °C (hold for 3 min); 5 °C/min to 290 °C (hold for 20 min) | ||
Interface/transfer line temperature: 280 °C (ion trap 200 °C) | Interface/transfer line temperature: 270 °C | ||||
MS source temperature: 230 °C | MS source temperature: 230 °C | ||||
Dwell time: | 50 ms | ||||
Ionization mode: | Electron ionization | ||||
Detection: | Full-scan m/z 30-300 | Ion peak for identification (m/z): | |||
Ion peak for identification(m/z): | Quantification | Confirmation | |||
B[ |
252 | B[ |
252 | 126 | |
B[ |
264 | B[ |
264 | 132 |
SOP 09. Standard Operating Procedure for Determination of Volatile Organics in Mainstream Cigarette Smoke under ISO and Intense Smoking Conditions 31.08.2018 ( |
CRM No. 70. Determination of Selected Volatile Organic Compounds in Mainstream Cigarette Smoke by GC-MS - September 2019 ( |
|||
---|---|---|---|---|
1,3-Butadiene, benzene | 1,3-Butadiene, isoprene, acrylonitrile, benzene, toluene | |||
Benzene- |
Benzene- |
|||
Mainstream cigarette smoke according to the Health Canada Intense (HCI) or the ISO regime is performed. One cigarette per smoke trap is smoked, containing 300 mg CX-572 particles and a glass fibre filter pad. The volatile organic components are extracted by adding a solution containing a mixture of carbon disulfide and methanol onto the Carboxen particles and glass fibre filter pad. The analytical determination of VOCs (1,3-butadiene and benzene) is performed by GC/MS with electron ionization mode. | VOCs are collected by passing the mainstream smoke of cigarettes through a GFP as specified in ISO 3308 or in ISO 20778 into cryogenic traps (impingers) containing methanol. The impinger solutions are fortified with benzene- |
|||
Linear smoking machine or Rotary smoking machine GC/MS | Linear smoking machine or Rotary smoking machine, Dewar flasks, Two impingers connected in series, GC/MS | |||
Option A: Carbon disulfide and methanol | 10 mL of methanol is added to each impinger and placed into the Dewar flasks containing the dry ice / isopropanol solution (the temperature must be at or below −70 °C). | |||
Option B: Carbon disulfide and methanol (1:4) mixed solution, containing internal standard solution | ||||
Option A – Linear smoking machine | 1,3-Butadiene: | 5 – 50 μg/mL | ||
• 1,3-Butadiene: | 20 – 160 μg/mL | Isoprene: | 12 – 600 μg/mL | |
• Benzene: | 8 – 64 μg/mL | Acrylonitrile: | 4 – 200 μg/mL | |
Option A – Rotary smoking machine | Benzene: | 4 – 200 μg/mL | ||
• 1,3-Butadiene: | 10 – 80 μg/mL | Toluene: | 4 – 200 μg/mL | |
• Benzene: | 4 – 32 μg/mL | |||
Option B – Linear smoking machine | ||||
• 1,3-Butadiene: | 20 – 160 μg/mL | |||
• Benzene: | 8 – 64 μg/mL | |||
Option B – Rotary smoking machine | ||||
• 1,3-Butadiene: | 10 – 100 μg/mL | |||
• Benzene: | 4 – 40 μg/mL | |||
ISO smoking regime | ISO smoking regime | |||
• Linear smoking machine: 3 | • Linear smoking machine: 5 or 10 | |||
• Rotary smoking machine: 3 | • Rotary smoking machine: 10 | |||
Intense smoking regime | Intense smoking regime | |||
• Linear smoking machine: 3 | • Linear smoking machine: 3 or 6 | |||
• Rotary smoking machine: 3 | • Rotary smoking machine: 5 | |||
After all samples are smoked following ISO regime, each impinger is spiked with 100 μL of benzene- |
||||
GC/MS | GC/MS | |||
GC Column: | InertCap AQUATIC-2 with a 60-m column with 0.25-mm internal diameter and 1.4-μm film thickness is suitable. | GC Column: | Fused silica capillary column with a 60-m column with 0.25-mm internal diameter and 1.4-μm film thickness is suitable. | |
Injector temperature: | 240 °C (ion trap: 200 °C) | Injector temperature: | 150 °C | |
Injection mode: | Split | Injection mode: | Split | |
Injection split ratio: | 1:10 | Injection split ratio: | 1:30 | |
Injection volume: | 1 μL or 2 μL (Ion trap 1:5; after 1.5 min 1:120) | Injection volume: | 3 μL | |
Mode: | Constant flow | Injection split flow: | 30 mL/min | |
Flow rate: | 1.0 mL/min (ion trap 1.0 mL/min) | Column temperature: | 40 °C (6 min); 20 °C / min to 225 °C (hold for 6 min) | |
Column temperature: | 40 °C (6 min); 6 °C/min to 250 °C | |||
Interface/transfer line temperature: 180 °C (ion trap 200 °C) | Interface/transfer line temperature: 240 °C | |||
MS source temperature: 200 °C (ion trap 180 °C) | MS source temperature: | 240 °C | ||
Dwell time: | 50 ms | Acquisition mode: | SIM (or scan) | |
Ionization mode: | Electron ionization (ionizing voltage at 70 eV) | Solvent delay: | Column dependent | |
Detection: | Full-scan m/z 30 – 300 | Detection: | Full-scan m/z 40 – 200 | |
Ion peak for identification (m/z): | Ion peak for identification (m/z): | |||
Benzene: | 78 | Quantification Confirmation | ||
Benzene- |
84 | 1,3-butadiene | 54 | 53 |
1,3-Butadiene: | 54 | Isoprene | 67 | 68 |
Acrylonitrile | 52 | 53 | ||
Benzene | 78 | 77 | ||
Benzene- |
84 | 83 | ||
Toluene | 91 | 92 |
Parameters | Smoking regime |
||
---|---|---|---|
Standard | Intensive | ||
ISO | Massachusetts | Canada | |
Puff volume (mL) | 35 | 45 | 55 |
Puff duration (s) | 2 | 2 | 2 |
Puff frequency (/min) | 1 | 2 | 2 |
SOP 03. Standard Operating Procedure for Determination of Tobacco-Specific Nitrosamines in Mainstream Cigarette Smoke Under ISO and Intense Smoking Conditions – June 2014 ( |
CRM No. 63. Determination of tobacco specific nitrosamines in cigarette mainstream smoke – GC-TEA method, January 2019 ( |
|||
---|---|---|---|---|
NNN, NNK, NAT, and NAB | NNN, NAB, NAT, and NNK | |||
Deuterium-labelled 3-(1-nitrosopyrrolidin-2-yl)pyridine(NNN- |
||||
Deuterium-labelled 4-(methylnitrosamino)-1-(3-pyridyl)-1-butanone (NNK- |
||||
Deuterium-labelled |
||||
Deuterium-labelled |
||||
Conditioned cigarettes are smoked using standard procedures. Mainstream smoke is trapped on a glass fiber filter pad. After addition of an internal standard, the filter pad is extracted with ammonium acetate. The extract is filtered and analysed by high-performance liquid chromatography–tandem mass spectrometry (HPLC/MS-MS) with electrospray ionization. Analyte ions are detected in the MS-MS mode. | Conditioned cigarettes are smoked using standard procedures. Mainstream smoke is trapped on a glass fiber filter pad. After addition of an internal standard, the filter pad is extracted with dichloromethane. Sample clean-up of the extraction solution is accomplished with one of the following methods: | |||
Elution of the extract through an alumina column, followed by the elution of the TSNAs with an acetone/dichloromethane (50:50 Elution of the extract through a combined silica-alumina column, followed by the elution of the TSNAs with an 8% methanol in dichloromethane solution. |
||||
The extract is concentrated followed by quantitative analysis using gas chromatography with a thermal energy analyser for detection (GC-TEA). | ||||
Linear smoking machine or Rotary smoking machine | Linear smoking machine or Rotary smoking machine | |||
• HPLC system interfaced to MS-MS (triple quadrupole mass spectrometer) | • Gas chromatography with a thermal energy analyser | |||
Ammonium acetate | Dichloromethane | |||
NNN: | 0.5 – 200 ng/mL | NNN: | 50 – 2000 ng/mL | |
NAB: | 0.5 – 200 ng/mL | NAB: | 50 – 2000 ng/mL | |
NAT: | 0.5 – 200 ng/mL | NAT: | 50 – 2000 ng/mL | |
NNK: | 0.5 – 200 ng/mL | NNK: | 50 – 2000 ng/mL | |
ISO smoking regime | ISO smoking regime | |||
• Linear smoking machine: 5 | • Linear smoking machine: 10 | |||
• Rotary smoking machine: 20 | • Rotary smoking machine: 20 | |||
Intense smoking regime | ||||
• Linear smoking machine: 3 | ||||
• Rotary smoking machine: 10 | ||||
Extraction of filter pads | Extraction of filter pads | |||
The pads from the holders are removed, transferred into the 60-mL amber bottle and added 100 mmol/L aqueous ammonium acetate extraction solution (20 mL for 44-mm pads, 50 mL for 93-mm pads) + internal standard (200 μL for a 44-mm pad, 500 μL for a 92-mm pad). The extraction of nitrosamines from the pad is for 30 min. Remove aliquots (e.g. 1–2 mL) of the extract, filter the aliquots with at most 0.45 μm membrane filter, and place in an autosampler vial. | The pads from the holders are removed, transferred into the Erlenmeyer flask and added 100 mL dichloromethane + 400 μL internal standard (10 cigarettes) or 200 mL dichloromethane + 800 μL internal standard (20 cigarettes). The extraction of nitrosamines from the pad is for 30 min. The volume is completed with dichloromethane to 150 mL (10 cigarettes) and 300 mL (20 cigarettes). An aliquot of the solution is concentrated to approximately 5 mL. The elution solution is concentrated to a volume 2 mL. | |||
HPLC system interfaced to MS-MS | Gas chromatograph – Thermal Energy Analyser (GC-TEA) | |||
HPLC column capable of distinct separation of TSNA and isotope labelled TSNA peaks from those of other cigarette emission components, e.g. Agilent Zorbax Eclipse XDB-C18 (2.1 x 150 mm, 3.5-μm particle size) | GC Column: Fused silica capillary column with a 50 % methyl/50 % phenyl polysiloxane stationary phase, a 30-m column with a 0.53mm internal diameter and 1-μm film thickness is suitable for this analysis. | |||
Injection volume: | 5 μL | Injector temperature: | 230 °C | |
Column oven temperature: | 40 ± 1 °C (Alter as appropriate for the column used.) | Injection mode: | Splitless | |
Injection volume: | 2 μL | |||
Degasser: | On | Mode: | Constant flow | |
Binary pump flow: | 0.2 mL/min | Column temperature: | 150 °C for 2 min, 3 °C/min to 230 °C, 20 °C/min to 250 °C, hold at 250 °C for 3 min | |
Mobile phase A: | 0.1 % acetic acid in water ( |
|||
Mobile phase B: | 0.1 % acetic acid in methanol ( |
|||
Gradient: | Shown in |
|||
Total analysis time: | 12 min | TEA conditions | ||
Mobile Phase A: | 0.1% acetic acid in water ( |
TEA interface temperature: 240 °C | ||
Mobile Phase B: | 0.1% acetic acid in methanol ( |
TEA pyrolysis temperature: 500 °C | ||
Gradient conditions: | ||||
0.0 | 50 % A | 50 % B | ||
3.0 | 10 % A | 90 % B | ||
4.0 | 0 % A | 100 % B | ||
5.0 | 0 % A | 100 % B | ||
5.5 | 50 % A | 50 % B | ||
12.0 | 50 % A | 50 % B |