Offering a systematic and multivariate analysis of the analytical procedure, development and validation of HPLC methods using Quality by Design approach are in the limelight of current research trends. A new, experimental design-aided HPLC method for fampridine was developed and preliminarily validated according to current in-force international guidelines for linearity, accuracy, robustness and precision.
The method offers a high throughput sample analysis, with an elution time of 2.9 minutes, and signal detection without excipient interference performed at 262 nm. The method proved to be linear between 1–15 µg mL−1 (