Evaluation of the visibility of peri-implant bone defects using ultrasonography with two types of probes


Cite

Fig. 1.

A. Image using both zirconium and titanium implants with fenestrations, B. Fenestration defects around titanium implants, C. Preparation of scanning site; note that the flap was again placed on the scanned region, D. Intra-oral scan, E. Extra-oral scanning of the dental implant sites
A. Image using both zirconium and titanium implants with fenestrations, B. Fenestration defects around titanium implants, C. Preparation of scanning site; note that the flap was again placed on the scanned region, D. Intra-oral scan, E. Extra-oral scanning of the dental implant sites

Fig. 2.

A. US image showing fenestration defects around titanium implants (arrows), B. US showing a fenestration defect around zirconium implant (arrow) scanned with a 13 MHz intra-operative probe that was used intraorally
A. US image showing fenestration defects around titanium implants (arrows), B. US showing a fenestration defect around zirconium implant (arrow) scanned with a 13 MHz intra-operative probe that was used intraorally

Fig. 3.

A. US image showing a fenestration defect around titanium implant (arrowhead), while the fenestration defects were non-visible around A. titanium and B. zirconium implant scanned with a 5–13.3 MHz linear probe that was used extraorally (arrows)
A. US image showing a fenestration defect around titanium implant (arrowhead), while the fenestration defects were non-visible around A. titanium and B. zirconium implant scanned with a 5–13.3 MHz linear probe that was used extraorally (arrows)

Fig. 4.

A. US image showing fenestrations around titanium (left arrow) and zirconium implant (right arrow) scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally without any fenestration defect visible
A. US image showing fenestrations around titanium (left arrow) and zirconium implant (right arrow) scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally without any fenestration defect visible

Fig. 5.

A. US image showing fenestrations around titanium implants scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally with the defect
A. US image showing fenestrations around titanium implants scanned with a 14 MHz intra-operative probe, B. Same implants scanned with a 4–9 MHz linear probe that was used extraorally with the defect

Interobserver agreement between observers. p value less than 0.05 considered as statically significant (95% CI)

ICC p
1st hockey stick probe intraorally 0.784 (0.349-0.930) 0.002
2nd hockey stick probe intraorally 0.909 (0.721-0.971) 0.001
1st linear probe extraorally 0.762 (0.300-0.922) 0.006
2nd linear probe extraorally 0.914 (0.728-0.972) 0.001

Intraobserver agreement for Observer 1 and Observer 2. p value less than 0.05 considered as statically significant (95% CI)

Observer 1 Observer 2
ICC p ICC p
1st hockey stick probe intraorally 0.878(0.632-0.960) 0.0001 0.952(0.817-0.986) 0.001
2nd hockey stick probe intraorally 0.966(0.885-0.989) 0.0001 0.921(0.752-0.975) 0.001
1st linear probe extraorally 0.696(0.095-0.901) 0.019 0.727(0.120-0.913) 0.016
2nd linear probe extraorally 0.849(0.524-0.952) 0.001 0.875(0.609-0.960) 0.001

Comparison of observers with the gold standard. p value less than 0.05 considered as statically significant

Sensitivity Specificity PPV NPV κ p
1st hockey stick probe intraorally Observer 1-Gold Standard 80% 100% 100% 90% 0.837 0.001
2nd hockey stick probe intraorally Observer 1-Gold Standard 80% 89% 80% 89%hz 0.689 0.011
1st linear probe extraorally Observer 1-Gold Standard 100% 33% 46% 100% 0.263 0.145
2nd linear probe extraorally Observer 1-Gold Standard 100% 33% 46% 100% 0.263 0.145
1st hockey stick probe intraorally Observer 2-Gold Standard 100% 89% 83% 100% 0.851 0.001
2nd hockey stick probe intraorally Observer 2-Gold Standard 100% 78% 71% 100% 0.714 0.005
1st linear probe extraorally Observer 2-Gold Standard 60% 22% 30% 50% –0.145 0.481
2nd linear probe extraorally Observer 2-Gold Standard 80% 11% 33% 50% –0.068 0.649