Charged projectile spectrometry using solid-state nuclear track detector of the PM-355 type
Publicado en línea: 25 sept 2015
Páginas: 591 - 596
Recibido: 21 oct 2014
Aceptado: 20 may 2015
DOI: https://doi.org/10.1515/nuka-2015-0100
Palabras clave
© by Aneta Malinowska
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
To use effectively any radiation detector in high-temperature plasma experiments, it must have a lot of benefits and fulfill a number of requirements. The most important are: a high energy resolution, linearity over a wide range of recorded particle energy, high detection efficiency for these particles, a long lifetime and resistance to harsh conditions existing in plasma experiments and so on. Solid-state nuclear track detectors have been used in our laboratory in plasma experiments for many years, but recently we have made an attempt to use these detectors in spectroscopic measurements performed on some plasma facilities. This paper presents a method that we used to elaborate etched track diameters to evaluate the incident projectile energy magnitude. The method is based on the data obtained from a semiautomatic track scanning system that selects tracks according to two parameters, track diameter and its mean gray level.