Electrical impedance tomography (EIT) is an imaging technique to image the impedance changes of the human body. It produces cross-sectional images of a conductivity distribution inside the human body using measured boundary voltage data sets which typically use a single ring external electrode. EIT has been widely used for functional imaging associated with the different physiological changes such as cardiac and respiratory changes. The first working device of EIT was presented by Brown
Fig. 1
Basic EIT arrangement.

Electrical impedance tomography (EIT) is a non-invasive, inexpensive and portable imaging modality which has a promising future to be used for bedside monitoring [7,8]. Numerous attempts have been taken earlier to extract cardiac and respiration signals properly. However those methods have their own limitations and are not satisfactory to obtain proper independent signal of cardiac and respiration signal. Consequently Independent Component Analysis (ICA) is a way to resolve signals into independent components based on statistical characteristics of the signals [9]. This method has been successfully pertained in EIT before. The EIT recordings are the result of linear combinations of the source signals. Using ICA the signal subspace will then be decomposed into statistically independent components [9]. At the present it is high time to apply this method in some extended clinical applications of EIT.
To reconstruct images from EIT datasets it is necessary to get all the individual conductivity distribution of the source signal. ICA is a method for factoring probability densities of measured signals into a set of densities that are as statistically independent as possible under the assumptions of a linear model. The EIT datasets are the result of linear combinations of the source signals. Using ICA the signal subspace we will decompose the linearly combined source signals into statistically independent components. Independent component analysis (ICA) is one of the most widely used Blind Source Separation (BSS) techniques for revealing hidden factors that underlie sets of random signals. It extracts individual signals from mixtures shown in Fig. 2.
Fig. 2
Independent Component Analysis block diagram where blind source

In this paper we want to propose a multi level spatial ICA based method that has been applied to acquire separate independent cardiac and respiration signal properly. This is the first ever attempt to extend the EIT based applications in the bed side monitoring. Firstly ICA gives the template function of most prominent respiration signal and again in the second level approximation ICA is been used to get another template function of cardiac signal from the time domain filtered signal. Then we analyze the template signals by using Fast Fourier Transform. Both of these template functions are later used in a reconstruction algorithm to get the impedance image. Also the computational time of multi-level spatial ICA is found reasonable to use this technique in real time system for bed side monitoring.
To study the previous relevant research, we organize our background study in three different existing approaches; ECG Gated Acquisition, Frequency Domain Filtering and recent PCA Based Methods.
ECG gated method is the first approach of separation of cardiac and respiration changes in electrical impedance tomography by McArdle
These methods are based on the fact that the heart rate and the respiratory rate are very well separated in frequency. The fast Fourier transform (FFT) of the global signal can provide a good approximation of the required cut-off frequency for filtering as shown in Fig. 3 showed by Grant
Fig. 3
(a) The original time course of impedance change of a subject during spontaneous breathing with no filtering applied. (b) The Fast Fourier Transform (FFT) power spectrum of this signal showing the frequency characteristics [15].

In spontaneously breathing subjects, the respiratory impedance changes are rarely sinusoidal, leading to frequency components across a wider range of spectrum than anticipated. Frerichs
A Principal Component Analysis (PCA) - based method for separating the respiratory and the perfusion-related signal was suggested in [14]. Their method successfully separated the two signals, allowing continuous beat-to-beat updates after a 20 s period of initiation. Time domain filtering with template functions was used to initially separate the cardiac and the ventilation signal. The results were then combined to obtain the best fit with the original signal, computed by performing least mean square (LMS) between the approximation and the actual signal.
In order to achieve proper separation of the respiratory and cardiac-related signals, the template functions for time domain filtering were identified by performing the PCA on the original data matrix. In [18] each of the aforementioned techniques compares the results achieved by each of these approaches in their experiment. Their results showed that the PCA and the frequency domain filtering methods produced highly correlated results. On the other hand, the correlation between these two methods with the ECG-gating alternative was reported considerably less. However, the correlations between each of these methods and a known standard method for perfusion imaging were not computed, rendering the results from their study inconclusive regarding as to which method is more reliable.
Impedance change of cardiac signal mixes with the impedance change of respiration signal and impedance change of blood flow. The EIT signal gets demodulated with the respiration signal of the human body and the cardiac signal gets demodulated too. As a result the output demodulated signal remains as the mixture of the cardiac and respiration signal together shown in Fig. 4. Based on our proposed method we are towards to put a novel impact in this inconveniency where we are to get the independent cardiac and respiration signal simultaneously. In the following sections our proposed method is described in detail.
Fig. 4
Separation of Cardiac and Respiration signal

In Fig. 5, we have shown the block diagram of our proposed method steps. Firstly ICA gives the template function of the most prominent respiration signal and again in the second level approximation ICA has been used to get another template function of cardiac signal from the time domain filtered signal. Then we analyze the template signals by using Fast Fourier Transform. Both of these template functions are later used in reconstruction algorithm to get the impedance image.
Fig. 5
Proposed method to Separate Cardiac and Respiration signal; After preprocessing ICA is applied on observed signal

In electrical impedance tomography, we preprocess the acquired data to decompose them into several independent activation maps. The source for each activation map will then be localized independently [9]. To apply ICA algorithm in EIT, we have made some assumptions [9, 19, 20, 21]. At First, the sources must be independent and is valid for our multiple conductivity distribution source of localization problem. The mixture is linear. There is no delay in signal propagation from sources to sensors and finally, the total number of independent signal sources does not exceed the number of electrodes.
To apply ICA in EIT we need to do some preprocessing [9]. A useful preprocessing technique in ICA is to whiten the observed signals. Whiten refers such transformation to preliminary sphere the data where the observed vector
The second step in processing the raw EIT data
The projection of the data on the signal subspace will be referred to as
Where
inverts the mixing process. M is unknown and the only data we have is the
where the vector
and for the nonlinear function
In the equation (3), α
The second phase of the ICA algorithm is the actual source separation. Independent components can be computed by applying the
Projection of independent activation maps
where
Experiment was done on a volunteered human subject. The data were collected from a normal male volunteer, without records of previous lung disease or smoking habits, and presenting a normal radiograph. The volunteer was upright during data collection, quietly breathing, with the electrodes placed at the level of the thoracic plane crossing the fifth intercostal space. Data was recorded using 16 channel EIT device for spontaneous breathing for continuous 40s.
In order to test the proposed method, data was collected from a human subject. EIT data were acquired using 16 channels KHU Mark2.5 with 16 electrodes. The electrodes were placed circumferentially (equally spaced) around the thorax just below the level of the axilla. An electrical current of 1mA was injected at 50 kHz through a pair of neighboring electrodes. Differential voltages were measured between the other non-injecting pair of electrodes with neighboring pattern. Following this initial current injection, the electrical current was then injected sequentially via the next pair of electrodes and repeated until all electrodes had served for current injection. The data for one complete cycle produced a so-called “frame,” and they were saved in a raw data file for later processing. The equipment has an SNR of 80 dB.
Considering all current patterns, 256 measurements were obtained. A 2-D finite element mesh with 499 linear triangular elements was built, with shape and size similar to the cross section of the volunteer’s thorax. Using the sensitivity matrix algorithm, 1650 measurement vectors were used to generate the transition matrix. Truncated singular valued decomposition (tSVD) algorithm was applied to get the reconstruction images.
In the experimental section, some intermediate results of our algorithm stages is shown and analyzed. For our experiment we first analyzed the input signals of our EIT data. We applied some preprocessing steps like whitening, PCA for denoising and to ensure the input data to be uncorrelated. Then we applied ICA to get the independent signals from our uncorrelated raw data. In Fig. 6 we have shown the sixteen independent components (IC’s) generated from ICA algorithm. A typical input signal as can be found in the lung region is depicted. The superposition of a lower frequency ventilation signal and a much smaller higher frequency cardiac signal can be distinguished. The signals are represented as frame numbers (horizontally) along with amplitude (vertically).
Fig 6
Independent Component signals generated from preprocessed raw EIT data.

All these IC’s are representing conductivity distribution up to some extent, which are verified by the Fast Fourier Transformation later on. Firstly, we applied FFT of all the selected IC’s generated from ICA algorithm. For our experiment, we have set the upper bound of the breathing rate which is 0.53 Hz (32 breath / min) with no modulation. The lower limit is estimated in our subjects able to breathe at rates of 0.4 Hz (24 breath / min) in most individuals. So, if we analyze the FFT of the IC’s, we will get high amplitude or peak in the region between 0.4 Hz to 0.52 Hz. From our analysis we analyze the IC’s by using FFT and observe the high amplitude of the signals. The high spectrum sustained on the above region is selected as candidate IC for our experiment. By analysis, we have selected IC_2, IC_6, IC_7 and IC_9 as our desired signal. Then we generate the reconstruction image of the IC’s and select the proper one for our desired respiration template RT. In Fig. 7 we have shown the reconstructed images of the selected IC’s. From the reconstruction images and from the rigorous survey we came to the conclusion that IC_9 represent the best respiration signal and we have chosen this independent respiration signal as RT and go further for the next step of the experiment. In Fig. 8 we have shown the magnified image of the RT and its FFT.
Fig. 7
Selected IC’s (left column) and corresponding reconstructed images (right column)

Fig. 8
Selected IC_9 with its magnified image (a) and its FFT (b).

For the next step we again use the preprocessing steps and the ICA algorithm as described in Fig. 5. After the first approximation and getting the RT, now we can get the remaining signal and get the cardiac signal as our desired template. By applying ICA on the remaining signal we can get the IC’s again. From the IC’s, we choose the right one as our desired template function CT. Then we generate the reconstruction image of the IC’s and select the proper one for our desired respiration template RT. In Fig. 9 we have shown the reconstructed images of the selected IC’s (right column). Also, the remaining IC’s which are selected and used for template function as cardiac template are shown in Fig. 10.
Fig. 9
Selected IC_3 with its magnified image and it’s reconstructed Image

Fig. 10
Selected IC’s from observed cardiac signals.

A new method for a dynamic extraction of cardiac and respiration changes in EIT image is proposed based on Independent Component Analysis. It does not rely on averaging over a number of heart cycles or on frequency domain filtering which is barely possible to implement properly. This method is reliable and further studies are going on to apply this method for different types of patients. Also the processing speed of this method is so fast that a continuous real-time monitoring of patients seems to be possible in the near future using this method.
Fig. 1

Fig. 2

Fig. 3
![(a) The original time course of impedance change of a subject during spontaneous breathing with no filtering applied. (b) The Fast Fourier Transform (FFT) power spectrum of this signal showing the frequency characteristics [15].](https://sciendo-parsed-data-feed.s3.eu-central-1.amazonaws.com/6005e020e797941b18f29fc9/j_jeb.553_fig_003.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Date=20230331T011018Z&X-Amz-SignedHeaders=host&X-Amz-Expires=18000&X-Amz-Credential=AKIA6AP2G7AKP25APDM2%2F20230331%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Signature=050b359da4fe7fd95e8d283a30d15411132210139aafeb653b84ba2abb3bf517)