Long Term Uncertainty Investigations of 1 MN Force Calibration Machine at NPL, India (NPLI)
Online veröffentlicht: 13. Aug. 2012
Seitenbereich: 149 - 152
DOI: https://doi.org/10.2478/v10048-012-0022-9
Schlüsselwörter
This content is open access.
The present paper is an attempt to study the long term uncertainty of 1 MN hydraulic multiplication system (HMS) force calibration machine (FCM) at the National Physical Laboratory, India (NPLI), which is used for calibration of the force measuring instruments in the range of 100 kN - 1 MN. The 1 MN HMS FCM was installed at NPLI in 1993 and was built on the principle of hydraulic amplifications of dead weights. The best measurement capability (BMC) of the machine is ± 0.025% (