Efficacy of breast shielding during head computed tomography examination
Artikel-Kategorie: Research Article
Online veröffentlicht: 01. Aug. 2020
Seitenbereich: 116 - 120
Eingereicht: 20. Mai 2020
Akzeptiert: 16. Juni 2020
DOI: https://doi.org/10.2478/raon-2020-0044
Schlüsselwörter
© 2021 Nika Zalokar, Nejc Mekis, published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
Background
Female breasts are exposed to scattered radiation regardless of not being included in the primary field during head CT. This study aimed to investigate whether the use of lead shielding is beneficial in dose reduction to the breasts during head CT.
Patients and methods
The study was performed in two different hospitals on two different CT units and included 120 patients. Half of the measurements (n = 60) was conducted without the use of lead shielding and the other half (n = 60) with the use of lead shielding of 0.5 mm equivalent thickness.
Results
Significant skin dose reduction to the breasts during head CT in both hospitals with the use of lead shielding was discovered; 81% (338.2 ± 43.7 μGy to 64.3 ± 18.8 μGy) in Hospital A and 74% (from 253.1 ± 35.1 μGy to 65.3 ± 16.9 μGy) in Hospital B.
Conclusions
Considering the assumed carcinogenic effect of low doses of radiation, high frequency of the head CT scans and the significant reduction of radiation doses to the highly radiosensitive breasts, the use of lead shielding is highly recommendable.