The effective and water-equivalent diameters as geometrical size functions for estimating CT dose in the thoracic, abdominal, and pelvic regions


Zitieren

Purpose: The aim of this work was to establish the relationships of patient size in terms of effective diameter (Deff) and water-equivalent diameter (Dw) with lateral (LAT) and anterior-posterior (AP) dimensions in order to predict the specific patient dose for thoracic, abdominal, and pelvic computed tomography (CT) examinations.

Methods: A total of 47 thoracic images, 79 abdominal images, and 50 pelvic images were analyzed in this study. The patient’s images were retrospectively collected from Dr. Kariadi and Kensaras Hospitals, Semarang, Indonesia. The slices measured were taken from the middle of the scan range. The calculations of patient sizes (LAT, AP, Deff, and Dw) were automatically performed by IndoseCT 20b software. Deff and Dw were plotted as functions of LAT, AP, and AP+LAT. In addition, Dw was plotted as a function of Deff.

Results: Strong correlations of Deff and Dw with LAT, AP, and AP+LAT were found. Stronger correlations were found in the Deff curves (R2 > 0.9) than in the Dw curves (R2 > 0.8). It was found that the average Deff was higher than the average Dw in the thoracic region, the average values were similar in the abdominal and pelvic regions.

Conclusion: The current study extended the study of the relationships between Deff and Dw and the basic geometric diameter LAT, AP, and AP+LAT beyond those previously reported by AAPM. We evaluated the relationships for three regions, i.e. thoracic, abdominal, and pelvic regions. Based on our findings, it was possible to estimate Deff and Dw from only the LAT or AP dimension.

eISSN:
1898-0309
Sprache:
Englisch
Zeitrahmen der Veröffentlichung:
4 Hefte pro Jahr
Fachgebiete der Zeitschrift:
Medizin, Biomedizinische Technik, Physik, Technische und angewandte Physik, Medizinische Physik