Electrical impedance characterization of in vivo porcine tissue using machine learning

Online veröffentlicht: 02 Jul 2021
Seitenbereich: 26 - 33
Eingereicht: 21 Mar 2021
Zeitschriftendaten
License
Format
Zeitschrift
Erstveröffentlichung
01 Jan 2010
Erscheinungsweise
1 Hefte pro Jahr
Sprachen
Englisch
Introduction

Surgical staplers are widely used in surgical procedures such as bariatric, thoracic, colorectal, and general surgery due to their versatility and efficiency (Fig. 1). However, each tissue type has its own challenges when stapling. For example, the stomach typically increases in thickness from the fundus down to the antrum and this must be accounted for when choosing the staple size to preserve staple line integrity and prevent leakage of gastric contents [1]. One of the drawbacks of endo-staplers that impacts stapling procedure is the loss of tactile feedback, which leads to a disconnect between surgeons and the tissue handling characteristics of these devices [2]. The incorporation of sensing technology within these instruments provides an opportunity for us to supplement, and even improve, our characterization of tissue handling in real time by enabling feedback control of the stapler. Multiple modalities of tissue characterization have been explored in the literature including, but not limited to optical [3], biomechanical [4], and bioelectric methods [5,6,7]. Electrical impedance has emerged as a promising tissue characterization modality as it offers good temporal resolution and has the advantages of being non-invasive and non-ionizing. However, it suffers from poor spatial resolution and is prone to artifacts from electrode movement and contact quality [8]. The utility of electrical impedance within medical instruments has already been demonstrated as advanced bipolar devices that deliver precise energy to seal and divide tissue by monitoring changes in tissue impedance have been on the market for over 20 years [9]. By advancing our understanding of tissue impedance further, we believe that bioimpedance can provide more sophisticated characterization beyond cauterized tissue to enable identification of tissue type and disease states.

Fig. 1

Medtronic's Endo GIA laparoscopic stapler

Electrical impedance spectroscopy (EIS) is a method used to generate complex impedance measurements of dielectric materials, by inducing alternating current over multiple frequencies to the material. EIS can be applied to biological tissue to determine its pathological and physiological state. Impedance of biological tissues is dependent on cellular features such as intracellular and cell membrane contents. Broad architectural features such as cellular size, density, spacing, and constituents of the extracellular matrix (ECM) also play an important role [10]. Intracellular and extracellular fluids are ionic solutions that provide conducting paths for alternating currents and contribute to the resistance portion of impedance. Cell membranes are a lipid bilayer and contribute to a capacitive reactance [10]. The electrical response of tissues has been divided into different segments of frequency responses known as dispersions regions [11]. The α dispersion encompasses lower frequencies up to several kHz and is related to electrical conduction of the extracellular fluid. The β dispersion region ranges from several kHz to tens of MHz and involves the cell membrane and conduction of intracellular fluid. The γ dispersion region from MHz to GHz is associated with biological macromolecules such as proteins and organelles [12]. Electrical impedance has already been studied extensively as a means to distinguish benign and cancerous tissue. Multiple studies have demonstrated significant differences in impedance between normal and malignant tissue for liver, breast, and prostate [13,14,15]. Since tissue bioimpedance is highly dependent on the water content within tissue, studies have also explored using impedance to identify areas of hemorrhage or monitor ischemia [16, 17]. Several groups have also developed custom impedance probes and demonstrated success in differentiating tissue types in both ex-vivo [6] and in-vivo [18] settings.

The traditional method of representing bioimpedance data has been with modeling. An equivalent circuit to model bioimpedance of tissue was produced by the work of Cole et al. [19] and has been widely adopted for bioelectric characterization [20,21,22]. The Cole function, shown below, contains four parameters to model bioimpedance data, Z(ω)=R+R0R1+(jωτ)α Z\left( \omega \right) = {R_\infty } + {{{R_0} - {R_\infty }} \over {1 + {{\left( {j\omega \tau } \right)}^\alpha }}} where ω is the angular frequency, R0 is the resistance at zero frequency, R is the resistance at infinite frequency, α is a dimensionless number between 0 and 1 that describes the shape of the curve, and τ is a time constant corresponding to the characteristic frequency fC (frequency at which the reactance is maximum) [13].

Machine learning (ML) in medicine has focused primarily on clinical imaging for feature extraction to streamline identification of cancer [23]. Improvements in algorithms and exponential growth in computing power have enabled analysis of large sets of high-dimensional data. While powerful, one of the major criticisms of ML is that machine-built models are often too complex for human understanding. The discrete relationships between variables can become muddled leading to a black box effect [24]. In general, machine learning (ML) has promise for tissue classification as increasingly large and complex data sets are generated. Expanded characterization efforts include not only simple tissue identification, but also disease states. Modeling methods have been effective in tissue impedance work when investigations have been limited in scope. However, when evaluating multiple tissue types and disease states, the complex pool of heterogenous data makes it increasingly difficult to correlate data to relevant equivalent circuit parameters, making ML more appealing and relevant. ML methods have been applied to bioimpedance data, albeit in a more limited scope. Several groups have applied machine learning algorithms to bioimpedance data in order to detect needle to nerve contact [25] and differentiate ischemic from healthy intestinal tissue in vivo [26]. But, to date, no one has performed classification of in vivo bioimpedance tissue data across multiple tissue types using ML methods.

In this work, we present and compare the results from in vivo electrical impedance measurements from pigs in order to attempt to use ML to classify healthy lung, liver, small bowel, colon, and stomach. To our knowledge, this is the first attempt at using ML to classify these in vivo tissue types based on EIS data.

Materials and methods

All impedance measurements were performed using a new, noncommercially available evaluation board (Analog Devices, Massachusetts, USA), based on but not identical to the AD5940. As per the manufacturer notes, the board utilizes a direct digital synthesis (DDS) chip that can set the test signal frequency in 0.2 Hz increments up to 10 MHz and change the amplitude with 16-bit resolution up to 2.4V. A customized flexible array strip (All Flex Inc) was designed with two rows 5 mm apart, each with ten gold leads 2.5 mm apart, mounted on a flexible copper strip. The sensor strip was interfaced with the module using a set of four co-axial cables. Impedance measurements in the literature have generally been performed with two or four electrode setups. For two electrode systems, the same pair of electrodes are used for both current injection and voltage measurement. As the alternating current is passed through the electrodes, contact impedance and electrode polarization gets added to the sample measurement which can lead to an over-estimation of tissue impedance [27]. While slightly more complex, the benefit of a four-electrode system is that measuring voltage with a very high input impedance prevents the flow of current in the sense electrodes [27]. Measurements in this study were performed using four-electrode testing arranged so that one pair of electrodes (outer electrodes) was used for current injection and another pair of electrodes (inner electrodes) was used for voltage measurement (see Fig. 2). This eliminates the electrode polarization impedance and allows the four-electrode system to perform more accurate measurements than two-electrode systems. All impedance measurements were obtained with a frequency sweep from 100 Hz to 1 MHz through fifty points, logarithmically. After each sweep, the measurements were exported as a data file. The data files were then exported and compiled into Excel using a custom MATLAB program.

Fig. 2

Electrode configuration. Zx represents board impedance measurement. H_CUR and L_CUR are connected to ground electrodes, while H_POT and L_POT are the sense electrodes.

The impedance measurement system was validated against multiple known resistors. The modulus and phase angle values obtained from a standard 10 kΩ resistor are depicted above in Fig. 3a and 3b. The modulus measurements show good accuracy with <1% error at the highest end of our testing frequency. The phase angle for an ideal resistor should be zero, but a linear dependency based on the frequency was identified in our system. There was a drift in the phase angle by -2E-5 degrees with each 1 Hz increase in frequency. At low frequencies, the phase angle remains low and essentially zero; however, this increases to a 20-degree difference at a frequency of 1 MHz. This difference can likely be attributed to capacitance within the measuring system. All subsequent measurements were adjusted accordingly to account for this phase angle frequency dependence.

Fig. 3

Validation of measurement system with known resistor values. (a) Modulus values of a control 10 kΩ resistor. (b) Phase angle values of a control 10 kΩ resistor. Phase angle measurements exhibit linear dependency on frequency with slope -2E-5.

In Vivo Porcine Testing

For the in vivo porcine tissue testing, a total of seven live pigs were used for impedance measurements. Healthy young domestic female pigs with weights ranging between 30–40 kg were utilized for testing. The pigs were sedated, intubated, and all animal procedures were conducted under a protocol approved by the Institutional Animal Care and Use Committee (IACUC) at Medtronic. The impedance testing was often performed as a secondary study, but no destructive testing was performed prior to impedance measurements. The hemodynamic status of the pig was continuously monitored, and testing was terminated if there was concern of hemodynamic compromise or tissue ischemia.

A midline laparotomy was initially performed. While pigs share similar metabolic and intestinal physiological processes with humans, and serve as good models for gastrointestinal tract, unlike humans, they have a larger and more developed cecum and the colon is arranged in a spiral structure [28]. The spiral colon was mobilized and attachments to itself were taken down to expose the colon for testing. The array strip (Fig. 4a) was manually placed against the tissue by an assistant ensuring that all four electrodes were in contact (Fig. 4b). The area of tissue examined by the electrode array is approximately 2 cm2. Impedance measurements were then obtained along different areas of the target tissue. The small bowel was then tested, followed by the liver. The attachments along the greater curve of the stomach were then taken down to mobilize the stomach for testing, making sure to maintain blood supply via the greater epiploic artery. Once all the intra-abdominal organs had been evaluated, a right thoracotomy was performed if feasible to provide exposure for lung testing. Due to a combination of lab availability constraints and animal instability, a thoracotomy was performed in only three pigs.

Fig. 4

Electrode placement. (a) is the electrode array that is described in section IIA, and shown in Fig. 2. Four total electrodes are used. The top and bottom white electrodes are H_CUR and L_CUR and are connected to ground electrodes, while the center yellow electrodes are H_POT and L_POT, the sense electrodes. (b) The electrode array is manually applied to the tissue for measurements.

Across all seven pigs, a total of 106 impedance measurements were taken on the stomach. Of these measurements, 16 were clearly designated on the antrum, 20 on the body, and 14 on the fundus, while the remaining 56 measurements were broadly labeled as stomach without further delineation. There were 121 unique measurements on the colon, 82 on the liver, and 108 on the small bowel. Lung measurements were only able to be performed on three different pigs for a total of 33 measurements.

Machine Learning

The validity of utilizing impedance for tissue differentiation was evaluated using the MATLAB classification learner machine learning tool. The resistance and reactance values for every tissue measurement were extracted from the data at seven frequencies: 145, 1400 11000, 49400, 184000, 570000, 1000000 for a total of fourteen parameters. The frequencies were arbitrarily chosen along a logarithmic scale based on previous experimental results to preserve the overall shape of the curve while also decreasing the processing burden. The models were run with 10-fold random cross-validation. Four methods were used to classify the data; these included (a) Decision Trees, (b) Support Vector Machines (SVM), (c) K-Nearest Neighbor Classifiers (KNN), and (d) Ensemble Classifiers. Each model was run ten times and the average accuracy and standard deviation was computed for each method and its associated classifier types. This was done to determine which method provided the most accurate characterization.

Details of the machine learning methods used in this study are available in detail from the MATLAB machine learning toolbox documentation [29], but are summarized here. Decision trees work by following the decisions in the tree from the root (starting) node as it splits down to a leaf node containing the response. Classifier types include coarse, medium, and fine which are determined by the maximum number of splits used (4, 20, and 100 respectively). As the number of splits increase, the likelihood of overfitting also increases. A fine tree with many leaves may be highly accurate on the training data but perform poorly when tested on an independent data set.

SVM classifies data by finding the hyperplane that optimizes the separation of data points between classes. The classifier type describes the shape of the hyperplane used to separate the data points. The classifier types for this method include linear, quadratic, cubic, and Gaussian which is further separated into fine, medium, and coarse. The kernel scale progressively decreases across the spectrum of coarse, medium, and fine Gaussian classifiers.

KNN works by categorizing a query point based on its distance to neighboring points in a training data set. The test point is assigned the label which is most frequent based on either a distance metric or set number of neighbors. Classifier types tested include fine, medium, coarse with increasing number of neighbors (1, 10, and 100, respectively). The classifiers set to 10 neighbors can be further subdivided into cosine, cubic, and weighted based on the structure of the distance metric. Nearest neighbor classifiers typically have good predictive accuracy in low dimensions but perform more poorly in higher dimensions.

Ensemble classifiers combine more than one machine learning algorithm into a predictive model. The classifier types are named based on the methods combined and include boosted trees, bagged trees, subspace discriminant, subspace KNN, and RUSBoosted trees.

Ethical approval

The research related to animal use complied with all the relevant national regulations and institutional policies for the care and use of animals.

Results

A Nyquist plot is a common way to present impedance data where the real resistance portion of the impedance is plotted on the X-axis against the imaginary reactance portion along the Y-axis. The average values of the aggregate data were used to generate the Nyquist plot for each tissue and the results are presented in Fig. 5a for all tissues. From this, it is clear that the liver and lung Nyquist curves have distinctly unique shapes, differentiating them from the colon, stomach, and small bowel. The colon, stomach, and small bowel have similar shapes, however, their position along the plot allows for differentiation.

Fig. 5

Nyquist plots for EIS measurements. (a) Comparison between all measured tissue types: colon, liver, small bowel, lung, and aggregate stomach. (b) Comparison between just stomach segments (fundus, body, antrum) and all stomach data taken as an aggregate.

During stomach impedance measurements, the specific segment (fundus, body, antrum) was clearly delineated for about half of the measurements. The glandular architecture is different within the segments of the stomach and there is an increase in thickness progressing from the fundus to body to antrum [30]. Due to the differences in composition, the impedance measurements of each segment would also be expected to vary. The variation between different organs was much larger than the difference between stomach segments. The Nyquist plots for each of the stomach segments is shown in Fig. 5b and can be compared to the aggregate stomach data. The shape of the stomach, antrum, fundus, and body plots are identical, but the antrum appears skewed to the right while the fundus is skewed to the left. The antrum is the thickest portion of the stomach while the fundus is thinner and floppier, and this may provide a physiological explanation for our observations.

There was a consistent change in character of the phase angle at approximately 50 kHz for all tissue types. This may be attributed to represent the shift between α and β dispersion regions; however, prior work performed by Strand-Amundsen et al. on the bioelectrical characterization of in vivo porcine small bowel did not visualize the same shift at 50 kHz [31]. Their group utilized a different impedance measurement system (Solartron). As this shift is maintained throughout all tissue samples, the disparity is postulated to be related to the testing setup. Veal et al. described how capacitance of the coaxial cables that connect the electrodes to the circuit can provide a high frequency conduction path to ground, so that some of the current that passes through the sample bypasses the electrometer producing inductive artifacts [32].

The data was parsed through the MATLAB classification learner three different ways: one where the stomach was segmented into body, fundus, antrum and compared to the small bowel, colon, liver, and lung; another where the stomach was taken as an aggregate; and finally, where the stomach was combined with the small bowel and colon to define GI tract structures and compared to the liver and lung. The mean accuracy and standard deviation along with the best performing subclass after applying the four different machine learning methods (decision trees, support vector machines, nearest neighbor classifiers, and ensemble classifiers) are presented in Fig. 6a. Support vector machines yielded the best results with accuracy of 86.96% when the stomach was segmented and 88.03% when stomach was taken as an aggregate. The antrum was the stomach segment that was most the difficult to identify with an accuracy of only 25% (Fig. 6b). When the fundus, body, and antrum are aggregated together as “stomach,” the stomach was the most frequently misclassified tissue, most often mistaken for the colon at 20% (Fig. 6c). When the stomach was combined with colon and small bowel to define the GI tract, all the algorithms had accuracy >98%, but fine KNN performed the best with 99.79% accuracy.

Fig. 6

(a) Comparison of mean accuracy with standard deviations for the different machine learning classification methods, along with best performing subclass. Confusion matrix plots generated from a single 10 fold cross-validation run using the cubic support vector machine subclass with (b) stomach separated out into fundus, body, and antrum and (c) fundus, body, and antrum aggregated together as stomach prior to comparison to other tissues.

Discussion

We recognize the limitation of this work due to the small dataset, but believe that the demonstration of this approach on actual tissue samples is important. Additionally, segmentation of the stomach divisions is limited compared to the small bowel and colon which have much larger surface areas for unique measurements. Future work is planned to test the classification ability of our model for unknown tissue samples from different pigs. Tissue properties have a major impact on stapling mechanics and technique. Suboptimal handling of tissue can increase the risk of complications such as anastomotic leaks. Lung tissue is uniquely challenging due to its heterogeneity. It is highly variable in the number of solid elements such as bronchi and vasculature and elastic elements such as alveoli [2]. These structural differences clearly differentiate lung tissue, and this was reflected in the observed bioimpedance profile. The stomach segments, colon, and small bowel all shared similarly shaped Nyquist curves and were positioned within the same vicinity on the plot. As a result, these tissues were commonly confused with each other during the ML classification. Within the segments of the stomach, the antrum was skewed in position compared to the fundus while maintaining the overall shape of the curve. It was difficult to differentiate the antrum from small bowel or colon, but it was never confused with the fundus. Identifying the various segments of stomach may provide clinical benefit in gastric resections as differences in gastric wall thickness have been postulated as a cause for staple line leaks [33]. The antrum, small bowel, and colon had comparable bioimpedance profiles which may be the result of similar architecture. Small bowel and colorectal tissue are much more uniform with similar stapling techniques for both tissues; however, if we can correlate tissue perfusion and tension with bioimpedance characteristics, modifications to current techniques can be made to reduce anastomotic leaks.

A benefit of impedance sensors is that they are typically lower profile compared to optical and mechanical sensors. This allows for easier incorporation onto pre-existing surgical instrumentation and stapling platforms. The electrode array strip used for this study was designed to provide flexibility with electrode arrangements and had the added benefit of fitting along the jaws of a surgical stapler. One of the short-comings of bioimpedance is that measurements are highly sensitive to both tissue inhomogeneity and electrode contact. The spacing between the electrodes affects the depth that the tissue is sampled, and further investigation remains to determine any discrepancies between superficial and transverse tissue measurements. Measurements within our study were obtained through manual application of the electrodes to the tissue surface and significant variation was seen between measurements even within the same tissue type. The amount of force and electrode-tissue contact was not standardized and compounded the variation between measurements. Additionally, the effect of compression on tissue impedance measurements remains poorly studied. Compression effects on impedance are different in ex vivo and in vivo settings, but the reason for this remains unclear [34]. Future work would focus on optimization of the testing setup to reduce contact variation as well as exploring the effects of tissue thickness and compression.

Conclusion

The focus of this study was to explore how impedance can be combined with ML to characterize different tissues. The results from our in vivo testing have expanded on bioimpedance characterization of multiple in vivo tissues and clearly demonstrate the capabilities of bioimpedance and ML for classifying tissue. There were obvious bioelectric differences observed between liver, lung, and gastrointestinal tissue with almost 100% classification accuracy across all machine learning algorithms. This work shows that impedance can be used to create a classification algorithm to identify commonly stapled tissue such as colon, small bowel, liver, stomach, and lung in vivo. With continued improvement through additional data that improves training of the ML tools, it can be possible to explore tissue compression effects and better control firing forces of surgical staplers depending on tissue type.

The feasibility of identifying tissue ischemia and malignancy has been proven repeatedly, but typically limited to a single tissue type, so next steps in this work would be to optimize the data collection and explore the feasibility of differentiating tissue states (healthy, ischemic, and malignant) using ML. Also, in vivo disease state, particularly for human tissues, remains limited. With more data collection, we can then develop tissue specific firing algorithms to optimize the stapling process such as site selection, staple height, and firing speed to improve surgical outcomes.

Fig. 1

Medtronic's Endo GIA laparoscopic stapler
Medtronic's Endo GIA laparoscopic stapler

Fig. 2

Electrode configuration. Zx represents board impedance measurement. H_CUR and L_CUR are connected to ground electrodes, while H_POT and L_POT are the sense electrodes.
Electrode configuration. Zx represents board impedance measurement. H_CUR and L_CUR are connected to ground electrodes, while H_POT and L_POT are the sense electrodes.

Fig. 3

Validation of measurement system with known resistor values. (a) Modulus values of a control 10 kΩ resistor. (b) Phase angle values of a control 10 kΩ resistor. Phase angle measurements exhibit linear dependency on frequency with slope -2E-5.
Validation of measurement system with known resistor values. (a) Modulus values of a control 10 kΩ resistor. (b) Phase angle values of a control 10 kΩ resistor. Phase angle measurements exhibit linear dependency on frequency with slope -2E-5.

Fig. 4

Electrode placement. (a) is the electrode array that is described in section IIA, and shown in Fig. 2. Four total electrodes are used. The top and bottom white electrodes are H_CUR and L_CUR and are connected to ground electrodes, while the center yellow electrodes are H_POT and L_POT, the sense electrodes. (b) The electrode array is manually applied to the tissue for measurements.
Electrode placement. (a) is the electrode array that is described in section IIA, and shown in Fig. 2. Four total electrodes are used. The top and bottom white electrodes are H_CUR and L_CUR and are connected to ground electrodes, while the center yellow electrodes are H_POT and L_POT, the sense electrodes. (b) The electrode array is manually applied to the tissue for measurements.

Fig. 5

Nyquist plots for EIS measurements. (a) Comparison between all measured tissue types: colon, liver, small bowel, lung, and aggregate stomach. (b) Comparison between just stomach segments (fundus, body, antrum) and all stomach data taken as an aggregate.
Nyquist plots for EIS measurements. (a) Comparison between all measured tissue types: colon, liver, small bowel, lung, and aggregate stomach. (b) Comparison between just stomach segments (fundus, body, antrum) and all stomach data taken as an aggregate.

Fig. 6

(a) Comparison of mean accuracy with standard deviations for the different machine learning classification methods, along with best performing subclass. Confusion matrix plots generated from a single 10 fold cross-validation run using the cubic support vector machine subclass with (b) stomach separated out into fundus, body, and antrum and (c) fundus, body, and antrum aggregated together as stomach prior to comparison to other tissues.
(a) Comparison of mean accuracy with standard deviations for the different machine learning classification methods, along with best performing subclass. Confusion matrix plots generated from a single 10 fold cross-validation run using the cubic support vector machine subclass with (b) stomach separated out into fundus, body, and antrum and (c) fundus, body, and antrum aggregated together as stomach prior to comparison to other tissues.

Susmallian S GD, Barnea R, Raziel A. Correct Evaluation of Gastric Wall Thickness May Support a Change in Staplers Size When Performing Sleeve Gastrectomy. The Israel Medical Association Journal: IMAJ. 2017;19:351–4. https://doi.org/10.1016/j.clnesp.2017.02.002 SusmallianS GD BarneaR RazielA Correct Evaluation of Gastric Wall Thickness May Support a Change in Staplers Size When Performing Sleeve Gastrectomy The Israel Medical Association Journal: IMAJ 2017 19 351 4 https://doi.org/10.1016/j.clnesp.2017.02.002 Search in Google Scholar

Chekan E, Whelan RL. Surgical stapling device-tissue interactions: what surgeons need to know to improve patient outcomes. Med Devices (Auckl). 2014;7:305–18. https://doi.org/10.2147/MDER.S67338 ChekanE WhelanRL Surgical stapling device-tissue interactions: what surgeons need to know to improve patient outcomes Med Devices (Auckl) 2014 7 305 18 https://doi.org/10.2147/MDER.S67338 Search in Google Scholar

Eriksson S, Nilsson J, Sturesson C. Non-invasive imaging of microcirculation: a technology review. Med Devices (Auckl). 2014;7:445–52. https://doi.org/10.2147/MDER.S51426 ErikssonS NilssonJ SturessonC Non-invasive imaging of microcirculation: a technology review Med Devices (Auckl) 2014 7 445 52 https://doi.org/10.2147/MDER.S51426 Search in Google Scholar

Baker RS, Foote J, Kemmeter P, Brady R, Vroegop T, Serveld M. The Science of Stapling and Leaks. Obesity Surgery. 2004;14(10):1290–8. https://doi.org/10.1381/0960892042583888 BakerRS FooteJ KemmeterP BradyR VroegopT ServeldM The Science of Stapling and Leaks Obesity Surgery 2004 14 10 1290 8 https://doi.org/10.1381/0960892042583888 Search in Google Scholar

Cheng Z, Dall'Alba D, Foti S, Mariani A, Chupin T, Caldwell DG, et al. Design and Integration of Electrical Bioimpedance Sensing in Surgical Robotic Tools for Tissue Identification and Display. Front Robot AI. 2019;6:55. https://doi.org/10.3389/frobt.2019.00055 ChengZ Dall'AlbaD FotiS MarianiA ChupinT CaldwellDG Design and Integration of Electrical Bioimpedance Sensing in Surgical Robotic Tools for Tissue Identification and Display Front Robot AI 2019 6 55 https://doi.org/10.3389/frobt.2019.00055 Search in Google Scholar

Rigaud B, Hamzaoui L, Frikha MR, Chauveau N, Morucci JP. In vitro tissue characterization and modelling using electrical impedance measurements in the 100 Hz-10 MHz frequency range. Physiological Measurement. 1995;16(3A):A15–A28. https://doi.org/10.1088/0967-3334/16/3A/002 RigaudB HamzaouiL FrikhaMR ChauveauN MorucciJP In vitro tissue characterization and modelling using electrical impedance measurements in the 100 Hz-10 MHz frequency range Physiological Measurement 1995 16 3A A15 A28 https://doi.org/10.1088/0967-3334/16/3A/002 Search in Google Scholar

Ruiz-Vargas A, Ivorra A, Arkwright JW. Design, Construction and Validation of an Electrical Impedance Probe with Contact Force and Temperature Sensors Suitable for in-vivo Measurements. Sci Rep. 2018;8(1):14818. https://doi.org/10.1038/s41598-018-33221-4 Ruiz-VargasA IvorraA ArkwrightJW Design, Construction and Validation of an Electrical Impedance Probe with Contact Force and Temperature Sensors Suitable for in-vivo Measurements Sci Rep 2018 8 1 14818 https://doi.org/10.1038/s41598-018-33221-4 Search in Google Scholar

Adler A, Boyle A. Electrical Impedance Tomography: Tissue Properties to Image Measures. IEEE Transactions on Biomedical Engineering. 2017;64(11):2494–504. https://doi.org/10.1109/TBME.2017.2728323 AdlerA BoyleA Electrical Impedance Tomography: Tissue Properties to Image Measures IEEE Transactions on Biomedical Engineering 2017 64 11 2494 504 https://doi.org/10.1109/TBME.2017.2728323 Search in Google Scholar

Karande VC. LigaSure™ 5-mm Blunt Tip Laparoscopic Instrument. J Obstet Gynaecol India. 2015;65(5):350–2. https://doi.org/10.1007/s13224-015-0745-2 KarandeVC LigaSure™ 5-mm Blunt Tip Laparoscopic Instrument J Obstet Gynaecol India 2015 65 5 350 2 https://doi.org/10.1007/s13224-015-0745-2 Search in Google Scholar

Bera TK. Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A Review. J Med Eng. 2014;2014:381251. https://doi.org/10.1155/2014/381251 BeraTK Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A Review J Med Eng 2014 2014 381251 https://doi.org/10.1155/2014/381251 Search in Google Scholar

Martinsen OG, Grimnes SG, Schwan HP. Interface phenomena and dielectric properties of biological tissue. Encycl Surf Colloid Sci. 2002:2643–52. MartinsenOG GrimnesSG SchwanHP Interface phenomena and dielectric properties of biological tissue Encycl Surf Colloid Sci 2002 2643 52 Search in Google Scholar

Ramírez-Chavarría RG, Sánchez-Pérez C, Matatagui D, Qureshi N, Pérez-García A, Hernández-Ruíz J. Ex-vivo biological tissue differentiation by the Distribution of Relaxation Times method applied to Electrical Impedance Spectroscopy. Electrochimica Acta. 2018;276:214–22. https://doi.org/10.1016/j.electacta.2018.04.167 Ramírez-ChavarríaRG Sánchez-PérezC MatataguiD QureshiN Pérez-GarcíaA Hernández-RuízJ Ex-vivo biological tissue differentiation by the Distribution of Relaxation Times method applied to Electrical Impedance Spectroscopy Electrochimica Acta 2018 276 214 22 https://doi.org/10.1016/j.electacta.2018.04.167 Search in Google Scholar

Gregory WD, Marx JJ, Gregory CW, Mikkelson WM, Tjoe JA, Shell J. The Cole relaxation frequency as a parameter to identify cancer in breast tissue. Medical Physics. 2012;39(7 Part1):4167–74. https://doi.org/10.1118/1.4725172 GregoryWD MarxJJ GregoryCW MikkelsonWM TjoeJA ShellJ The Cole relaxation frequency as a parameter to identify cancer in breast tissue Medical Physics 2012 39 7 Part1 4167 74 https://doi.org/10.1118/1.4725172 Search in Google Scholar

Halter RJ, Hartov A, Heaney JA, Paulsen KD, Schned AR. Electrical Impedance Spectroscopy of the Human Prostate. IEEE Transactions on Biomedical Engineering. 2007;54(7):1321–7. https://doi.org/10.1109/TBME.2007.897331 HalterRJ HartovA HeaneyJA PaulsenKD SchnedAR Electrical Impedance Spectroscopy of the Human Prostate IEEE Transactions on Biomedical Engineering 2007 54 7 1321 7 https://doi.org/10.1109/TBME.2007.897331 Search in Google Scholar

Laufer S, Ivorra A, Reuter VE, Rubinsky B, Solomon SB. Electrical impedance characterization of normal and cancerous human hepatic tissue. Physiological Measurement. 2010;31(7):995–1009. https://doi.org/10.1088/0967-3334/31/7/009 LauferS IvorraA ReuterVE RubinskyB SolomonSB Electrical impedance characterization of normal and cancerous human hepatic tissue Physiological Measurement 2010 31 7 995 1009 https://doi.org/10.1088/0967-3334/31/7/009 Search in Google Scholar

Dzwonczyk R, Rio Cd, Brown DA, Michler RE, Wolf RK, Howie MB. Myocardial electrical impedance responds to ischemia and reperfusion in humans. IEEE Transactions on Biomedical Engineering. 2004;51(12):2206–9. https://doi.org/10.1109/TBME.2004.834297 DzwonczykR RioCd BrownDA MichlerRE WolfRK HowieMB Myocardial electrical impedance responds to ischemia and reperfusion in humans IEEE Transactions on Biomedical Engineering 2004 51 12 2206 9 https://doi.org/10.1109/TBME.2004.834297 Search in Google Scholar

Yang L, Zhang G, Song J, Dai M, Xu C, Dong X, et al. Ex-Vivo Characterization of Bioimpedance Spectroscopy of Normal, Ischemic and Hemorrhagic Rabbit Brain Tissue at Frequencies from 10 Hz to 1 MHz. Sensors (Basel). 2016;16(11):1942. https://doi.org/10.3390/s16111942 YangL ZhangG SongJ DaiM XuC DongX Ex-Vivo Characterization of Bioimpedance Spectroscopy of Normal, Ischemic and Hemorrhagic Rabbit Brain Tissue at Frequencies from 10 Hz to 1 MHz Sensors (Basel) 2016 16 11 1942 https://doi.org/10.3390/s16111942 Search in Google Scholar

Dai Y, Du J, Yang Q, Zhang J. Noninvasive electrical impedance sensor for in vivo tissue discrimination at radio frequencies. Bioelectromagnetics. 2014;35(6):385–95. https://doi.org/10.1002/bem.21854 DaiY DuJ YangQ ZhangJ Noninvasive electrical impedance sensor for in vivo tissue discrimination at radio frequencies Bioelectromagnetics 2014 35 6 385 95 https://doi.org/10.1002/bem.21854 Search in Google Scholar

Cole K, Curtis H. Electrical physiology: Electrical resistance and impedance of cells and tissues, in Medical Physics. New York: Year Book Publishers; 1944. ColeK CurtisH Electrical physiology: Electrical resistance and impedance of cells and tissues, in Medical Physics New York Year Book Publishers 1944 Search in Google Scholar

Gholami-Boroujeny S, Bolic M. Extraction of Cole parameters from the electrical bioimpedance spectrum using stochastic optimization algorithms. Med Biol Eng Comput. 2016;54(4):643–51. https://doi.org/10.1007/s11517-015-1355-y Gholami-BoroujenyS BolicM Extraction of Cole parameters from the electrical bioimpedance spectrum using stochastic optimization algorithms Med Biol Eng Comput 2016 54 4 643 51 https://doi.org/10.1007/s11517-015-1355-y Search in Google Scholar

Seoane F, Buendia R, Gil-Pita R. Cole parameter estimation from electrical bioconductance spectroscopy measurements. Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:3495–8. https://doi.org/10.1109/IEMBS.2010.5627790 SeoaneF BuendiaR Gil-PitaR Cole parameter estimation from electrical bioconductance spectroscopy measurements Annu Int Conf IEEE Eng Med Biol Soc 2010 2010 3495 8 https://doi.org/10.1109/IEMBS.2010.5627790 Search in Google Scholar

Ayllon D, Seoane F, Gil-Pita R. Cole equation and parameter estimation from electrical bioimpedance spectroscopy measurements - A comparative study. Annu Int Conf IEEE Eng Med Biol Soc. 2009;2009:3779–82. https://doi.org/10.1109/IEMBS.2009.5334494 AyllonD SeoaneF Gil-PitaR Cole equation and parameter estimation from electrical bioimpedance spectroscopy measurements - A comparative study Annu Int Conf IEEE Eng Med Biol Soc 2009 2009 3779 82 https://doi.org/10.1109/IEMBS.2009.5334494 Search in Google Scholar

Bowen PK, Shearier ER, Zhao S, Guillory RJ, 2nd, Zhao F, Goldman J, et al. Biodegradable Metals for Cardiovascular Stents: from Clinical Concerns to Recent Zn-Alloys. Advanced healthcare materials. 2016;5(10):1121–40. https://doi.org/10.1002/adhm.201501019 BowenPK ShearierER ZhaoS GuilloryRJ2nd ZhaoF GoldmanJ Biodegradable Metals for Cardiovascular Stents: from Clinical Concerns to Recent Zn-Alloys Advanced healthcare materials 2016 5 10 1121 40 https://doi.org/10.1002/adhm.201501019 Search in Google Scholar

Schmidt J, Marques MRG, Botti S, Marques MAL. Recent advances and applications of machine learning in solid-state materials science. npj Computational Materials. 2019;5(1):83. https://doi.org/10.1038/s41524-019-0221-0 SchmidtJ MarquesMRG BottiS MarquesMAL Recent advances and applications of machine learning in solid-state materials science npj Computational Materials 2019 5 1 83 https://doi.org/10.1038/s41524-019-0221-0 Search in Google Scholar

Kalvoy H, Tronstad C, Ullensvang K, Steinfeldt T, Sauter AR. Detection of needle to nerve contact based on electric bioimpedance and machine learning methods. Conf Proc IEEE Eng Med Biol Soc. 2017;2017:9–12. https://doi.org/10.1109/EMBC.2017.8036750 KalvoyH TronstadC UllensvangK SteinfeldtT SauterAR Detection of needle to nerve contact based on electric bioimpedance and machine learning methods Conf Proc IEEE Eng Med Biol Soc 2017 2017 9 12 https://doi.org/10.1109/EMBC.2017.8036750 Search in Google Scholar

Strand-Amundsen RJ, Tronstad C, Reims HM, Reinholt FP, Høgetveit JO, Tønnessen TI. Machine learning for intraoperative prediction of viability in ischemic small intestine. Physiological Measurement. 2018;39(10):105011. https://doi.org/10.1088/1361-6579/aae0ea Strand-AmundsenRJ TronstadC ReimsHM ReinholtFP HøgetveitJO TønnessenTI Machine learning for intraoperative prediction of viability in ischemic small intestine Physiological Measurement 2018 39 10 105011 https://doi.org/10.1088/1361-6579/aae0ea Search in Google Scholar

Chowdhury A, Ghoshal D, Bera T, Chakraborty B, Naresh M. Comparison of two and four electrode methods for studying the impedance variation during cucumber storage using Electrical Impedance Spectroscopy (EIS). 2017. p. 261–5. https://doi.org/10.1201/9781315400624-50 ChowdhuryA GhoshalD BeraT ChakrabortyB NareshM Comparison of two and four electrode methods for studying the impedance variation during cucumber storage using Electrical Impedance Spectroscopy (EIS) 2017 261 5 https://doi.org/10.1201/9781315400624-50 Search in Google Scholar

Gonzalez LM, Moeser AJ, Blikslager AT. Porcine models of digestive disease: the future of large animal translational research. Transl Res. 2015;166(1):12–27. https://doi.org/10.1016/j.trsl.2015.01.004 GonzalezLM MoeserAJ BlikslagerAT Porcine models of digestive disease: the future of large animal translational research Transl Res 2015 166 1 12 27 https://doi.org/10.1016/j.trsl.2015.01.004 Search in Google Scholar

MathWorks. Machine Learning Toolbox [Available from: https://www.mathworks.com/help/stats/choose-a-classifier.html#bunt0ky. MathWorks Machine Learning Toolbox [Available from: https://www.mathworks.com/help/stats/choose-a-classifier.html#bunt0ky. Search in Google Scholar

Rawlins L, Rawlins MP, Teel D. Human tissue thickness measurements from excised sleeve gastrectomy specimens. Surgical Endoscopy. 2014;28(3):811–4. https://doi.org/10.1007/s00464-013-3264-1 RawlinsL RawlinsMP TeelD Human tissue thickness measurements from excised sleeve gastrectomy specimens Surgical Endoscopy 2014 28 3 811 4 https://doi.org/10.1007/s00464-013-3264-1 Search in Google Scholar

Strand-Amundsen RJ, Tronstad C, Kalvøy H, Gundersen Y, Krohn CD, Aasen AO, et al. In vivo characterization of ischemic small intestine using bioimpedance measurements. Physiological Measurement. 2016;37(2):257–75. https://doi.org/10.1088/0967-3334/37/2/257 Strand-AmundsenRJ TronstadC KalvøyH GundersenY KrohnCD AasenAO In vivo characterization of ischemic small intestine using bioimpedance measurements Physiological Measurement 2016 37 2 257 75 https://doi.org/10.1088/0967-3334/37/2/257 Search in Google Scholar

Veal B, Baldo P, Paulikas A, Eastman J. Understanding Artifacts in Impedance Spectroscopy. Journal of the Electrochemical Society. 2015;162:H47–H57. https://doi.org/10.1149/2.0791501jes VealB BaldoP PaulikasA EastmanJ Understanding Artifacts in Impedance Spectroscopy Journal of the Electrochemical Society 2015 162 H47 H57 https://doi.org/10.1149/2.0791501jes Search in Google Scholar

Barski K, Binda A, Kudlicka E, Jaworski P, Tarnowski W. Gastric wall thickness and stapling in laparoscopic sleeve gastrectomy - a literature review. Wideochir Inne Tech Maloinwazyjne. 2018;13(1):122–7. https://doi.org/10.5114/wiitm.2018.73362 BarskiK BindaA KudlickaE JaworskiP TarnowskiW Gastric wall thickness and stapling in laparoscopic sleeve gastrectomy - a literature review Wideochir Inne Tech Maloinwazyjne 2018 13 1 122 7 https://doi.org/10.5114/wiitm.2018.73362 Search in Google Scholar

Moqadam S, Grewal P, Shokoufi M, Golnaraghi M. Compression-dependency of soft tissue bioimpedance for in-vivo and in-vitro tissue testing. Journal of Electrical Bioimpedance. 2015;6:22–32. https://doi.org/10.5617/jeb.1489 MoqadamS GrewalP ShokoufiM GolnaraghiM Compression-dependency of soft tissue bioimpedance for in-vivo and in-vitro tissue testing Journal of Electrical Bioimpedance 2015 6 22 32 https://doi.org/10.5617/jeb.1489 Search in Google Scholar