Dielectrical properties of living epidermis and dermis in the frequency range from 1 kHz to 1 MHz
, , , und
02. Juli 2019
Über diesen Artikel
Online veröffentlicht: 02. Juli 2019
Seitenbereich: 14 - 23
Eingereicht: 28. Nov. 2018
DOI: https://doi.org/10.2478/joeb-2019-0003
Schlüsselwörter
© 2019 Tsai, Xue, Birgersson, Olmar, Birgersson, published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.
Fig.1

Fig. 2
![The mean (depth 1 ●; depth 5 ■] and standard deviations (error bars) for the (a) magnitude and (b) phase of the experimentally measured impedances. The fitted mean resistivity and relative permittivity and their standard deviations can be found in (c) and (d) respectively for the living epidermis (♦) and dermis (▼); the fitted equations, Eqs. 7-8, are shown as lines. In (e) and (f), the validation set for depth setting 3 is shown for the experiments (●) and model predictions (lines).](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/64721e3e215d2f6c89dbc8fe/j_joeb-2019-0003_fig_002.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Content-Sha256=UNSIGNED-PAYLOAD&X-Amz-Credential=AKIA6AP2G7AKOUXAVR44%2F20250911%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Date=20250911T112117Z&X-Amz-Expires=3600&X-Amz-Signature=c1bb8e251dcfd5dc3786fddebf0fe9020cf8c232aa39ff0a33fb3c88af754972&X-Amz-SignedHeaders=host&x-amz-checksum-mode=ENABLED&x-id=GetObject)
Skin thickness (mean value ± standard deviation) at the volar forearm for epidermis, stratum corneum and dermis_ Some measurement techniques are: Confocal Microscopy (CM), Confocal laser scanning microscopy (CLSM), Optical coherence tomography (OCT), Transepidermal water loss (TEWL), Confocal Raman spectrometer (CRS) and Ultrasound (US)_
ED (μm) | Subj | Age | Measurement | Source |
---|---|---|---|---|
73.54±2.84 | 10 | 23 - 47 | CM | [12] |
89±9 | 15 | 19 - 24 | CLSM | [13] |
71.8±10 | 30 | 21 - 35 | OCT | [14] |
102±7.4 | 5 | 20 - 35 | OCT | [15] |
66.39±5.82 | 19 | 20 - 29 | OCT | [16] |
65.1±8.9 | 8 | 31 - 37 | OCT | [17] |
Dimensions and material parameters_
Current detection width, | 1 mm |
Ceramic width, | 0.15, 0.15, 1.9 mm |
Guard width, | 0.3 mm |
Secondary inject width, | 0.5 mm |
Primary inject width, | 0.5 mm |
Living epidermis thickness, | 61 μm |
Dermis thickness, | 1.0 mm |
Inject voltage, | 0.05 V |
Depth setting, | 0.1, 0.58, 1 |
Electrical permittivity in vacuum, | 8.85 x10-12 Fm-1 |
Coefficients for the conductivity and relative permittivity of the skin layers for 1 kHz to 1 MHz_
0 | 1.6306×10-2 | -2.3922×10-2 |
1 | -3.3591×10-1 | 5.5919×10-1 |
2 | 2.7019×100 | -5.2033×100 |
3 | -1.0667×101 | 2.4090×101 |
4 | 2.0531×101 | -5.6056×101 |
5 | -1.3260×101 | 5.7328×101 |
0 | -1.0475×10-2 | -2.5399×10-1 |
1 | 2.2902×10-1 | 5.5554×100 |
2 | -1.9609×100 | -4.7903×101 |
3 | 8.2439×100 | 2.0330×102 |
4 | -1.7209×101 | -4.2527×102 |
5 | 1.4687×101 | 3.5741×102 |