Zitieren

Introduction: The subcutaneous implantable cardioverter defibrillator (S-ICD) represents a major advancement in ICD technology. Inappropriate shocks (IAS) occur in more than 3.1% of the population with S-ICD each year and are usually followed by admission to the emergency department (ED). In this setting, the disabling of IAS is mandatory during a pseudo-electrical storm (ES). This report describes the strategies that can be followed in order to temporarily inhibit IAS in critical care settings with the use of magnets.

Case presentation: An S-ICD was implanted more than 6 weeks prior to presentation in a 68-year-old man with hypertrophic cardiomyopathy. In the ED, the patient experienced 3 IAS in the last hour. A Medtronic magnet was applied to stop IAS, as the specific programmer was not available. The maneuver interrupted the IAS. In order to verify the universal magnetic response of the S-ICD, six different magnets and one smartphone with MagSafe technology were tested. All magnet models suspended arrhythmia detection and IAS, while the smartphone did not cause magnet interferences.

Conclusions: This report demonstrates the safety and efficacy of all clinical magnet models in inhibiting IAS. In case of pseudo-ES, any type of magnet allows ED providers to easily and rapidly disable the functionality of the devices when appropriate.

eISSN:
2457-5518
Sprache:
Englisch
Zeitrahmen der Veröffentlichung:
4 Hefte pro Jahr
Fachgebiete der Zeitschrift:
Medizin, Klinische Medizin, Intensivmedizin und Notfallmedizin, Radiologie, Allgemeinmedizin, Innere Medizin, Kardiologie