Measures of acromiohumeral distance with wireless ultrasound machine in subacromial impingement syndrome: an inter-machine reliability study
Artikel-Kategorie: Research paper
Online veröffentlicht: 08. Nov. 2024
Seitenbereich: 1 - 5
Eingereicht: 10. Juli 2023
Akzeptiert: 11. Jan. 2024
DOI: https://doi.org/10.15557/jou.2024.0018
Schlüsselwörter
© 2024 Burak Tayyip Dede et al., published by Sciendo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Shoulder pain is the third most common musculoskeletal cause of admissions for treatment(1). Subacromial impingement syndrome (SIS) is among the most typical causes of shoulder pain. SIS occurs when the rotator cuff tendons, especially the supraspinatus tendon (SST), are compressed between the acromion and the humerus(2). Many factors are thought to lead to the development of SIS. Extrinsic processes compressing the subacromial space and intrinsic changes in the supraspinatus tendon have both been shown to be contributing factors(3,4). Different imaging modalities have measured the acromiohumeral distance (AHD) in the literature to assess the subacromial space(5–7). However, the advantage of ultrasonography (US) among these imaging modalities is that it is fast, inexpensive, real-time, provides easy access, and causes no radiation exposure. In SIS, symptoms occur especially during arm lifting, and the US, which allows dynamic imaging, can reveal changes in the subacromial space during the activity of lifting(6,8).
McCreesh
The use of WUS devices is gradually increasing; however, it is necessary to compare these devices with SUS and investigate their reliability. This study was undertaken to evaluate the inter-machine reliability of AHD measurements in patients with SIS.
This study had a cross-sectional design. The study protocol was approved by the Istanbul Training and Research Hospital Ethics Committee (approval no: 2011-KAEK-50, Decision No. 61). A written informed consent was obtained from each patient. The study was conducted following the principles of the Declaration of Helsinki.
Patients presenting to our clinic with shoulder pain were screened for eligibility. The diagnosis of SIS was made by EA, a specialist with a 15-year interest in musculoskeletal disorders, in the presence of at least three of the five positive clinical tests (Neer, Hawkins-Kennedy, painful arch, Jobe, and external rotation resistance) on physical examination(16). The inclusion criteria were patients aged 18–55 years with unilateral shoulder pain lasting longer than six weeks. Patients were excluded from the study if they presented with bilateral shoulder pain, history of rheumatological diseases, diabetes mellitus, chronic liver or kidney failure, cervical pathologies, previous upper extremity/shoulder fracture or surgery, positive apprehension test, adhesive capsulitis (<90° of passive abduction and external rotation)(17), hook-shaped acromion and severe glenohumeral osteoarthritis evident, or a full-thickness tear on Magnetic Resonance Imaging (MRI). The study also excluded patients who had received steroid injections or physiotherapy in the shoulder area in the last six months, and those who had received anti-inflammatory-analgesic treatment over the past week.
Demographic data (age, gender, body mass index, dominant extremity) of the patients included in the study were recorded at the beginning of the study.
Two different linear probes at a frequency of 5–10 MHz were used on a mode B standard ultrasound device (Philips Affiniti 50, Medical International Co., Ltd., Amsterdam, The Netherlands) and a wireless ultrasound device (Clarius™ ultrasound scanner linear, L7, Clarius, Burnbaby, BC, Canada). FB, a specialist with at least five years of expertise in musculoskeletal ultrasonography, assessed the acromiohumeral distance. Before the start of the study, FB performed a pilot ultrasound examination with SUS and WUS in the neutral and 60° abduction positions on 10 patients (fulfilling the study criteria) until he was confident that reproducible measurements would be obtained. Patients who participated in the pilot examination were excluded from the study. In the ultrasound scanner, preset ultrasonographic settings were selected for musculoskeletal imaging, including depth, focal point, and gain. These settings were adjusted according to the ultrasound image and the person being scanned to ensure optimal visualization of the anatomical landmarks. To obtain images with the wireless probe, the Wireless USG application was downloaded from the app store of the tablet or mobile device. In our study, we used an iPad (screen size 20 cm ×15 cm) and provided a WI-FI connection to the iPad with the wireless probe’s password. The device became usable with a WI-FI connection, and real-time images were instantly transferred to the iPad. Each measurement was kept in the SUS database and on the iPad. After all scans of all study participants were completed, AHD measurements were performed three times at weekly intervals by FB, using two different devices for each patient. The measurements made by FB were recorded by BTD, and FB was prevented from accessing the measurement results again. The mean of the recorded measurements was calculated by BTD.
AHD measurements in the neutral position were performed with the patients seated in an upright posture in a chair, at rest, with their elbows in a 90° flexion position and their forearms resting on a pillow in their laps in the shoulder neutral position(18). The ultrasound transducer was placed longitudinally on the anterior surface of the anterior acromial border to visualize the humeral head and the acromion. The acromiohumeral distance was defined as the linear perpendicular distance between the upper surface of the humeral head and the lower surface of the acromion(12,19) (Fig. 1 A).

The shoulder was passively abducted to 60° with the help of the Orthoservice SAW 60° shoulder abduction brace. The patient was held in an upright position in a chair, and the elbow was flexed to 90°. The measurement was confirmed with a goniometer. The ultrasound transducer was placed longitudinally on the anterior surface of the anterior acromial border to visualize the humeral head and the acromion. The acromiohumeral distance was defined as the linear perpendicular distance between the upper surface of the humeral head and the lower surface of the acromion(19) (Fig. 1 B).
The sample size was determined using Power Analysis and Sample Size Software 15 (2017) (NCSS, LLC., Kaysville, UT, USA;
Statistical analysis was performed using IBM SPSS version 22.0 software (IBM Corp., Armonk, IL, USA). The normal distribution was evaluated by the Kolmogorov-Smirnov/Shapiro-Wilk test, kurtosis and skewness values, and histogram plots. Mean ± standard deviation and median (minimum–maximum) values were given when presenting descriptive analyses. Bland-Altman analysis was used to detect systematic bias by comparing the differences between machines according to their means with 95% limits of agreement (LOA)(20). Standard error of the measurement (SEM) and minimal detectable change (MDC) values were determined. The SEM value provides an estimate of how much samples differ within a population. MDC is the smallest difference detected that is clinically significant for the measurement to be considered reliable, i.e. the value at which differences are most likely not due to measurement error(10). SEM = (SD) × √(1-ICC) and MDC = (SEM) × (√2) × (1.96). Here, 1.96 represents a 95% confidence level. The inter-machine reliability of US measurements was assessed by calculating intraclass correlation coefficients (ICC) with 95% CI and absolute agreement for a single measurement. Statistical significance was determined as
A total of 61 participants with SIS, 30 (49.2%) males and 31 (50.2%) females, were included in our study. Demographic data of the participants are shown in Tab. 1. AHD_0 was measured as 11.59 ± 1.38 mm and 11.63 ± 1.34 mm for SUS and WUS, respectively. AHD_60 was found to be 6.34 ± 1.01 mm and 6.32 ± 1.02 mm for SUS and WUS, respectively. AHD_0 and AHD_60 were not significantly different between the two machines (

Patients characteristics
Mean ± SD | Median (min-max) | |
---|---|---|
Height (cm) | 163.84 ± 7.96 | 165 (149–180) |
Weight (kg) | 69.98 ± 12.43 | 70 (35–102) |
BMI (kg/cm2) | 25.7 ± 4.95 | 25 (15–37) |
Age (years) | 45.25 ± 8.63 | 48 (26–59) |
BMI – body mass index
Comparison of patients’ AHDs measured with different devices and at different angles
AHD | Standard US ( |
Wireless US ( |
|
---|---|---|---|
AHD_0 | 11.5 ± 1.38 | 11.6 ± 1.3 | 0.398 |
11.7 (8.8–16.3) | 11.5 (9–16.2) | ||
AHD_60 | 6.3 ± 1.0 | 6.3 ± 1.0 | 0.609 |
6.4 (3.3–8.9) | 6.5 (3.3–8.9) |
P* Paired Sample T-Test. Values are means ± standard deviation (SD) and median (minimum-maximum).
AHD – acromiohumeral distance; US – ultrasound;
Inter-machine reliability analysis
ICC | %95 CI | SEM (mm) | MDC (mm) | |
---|---|---|---|---|
AHD_0 | 0.97 | 0.95–0.98 | 0.23 | 0.63 |
AHD_60 | 0.96 | 0.94–0.97 | 0.20 | 0.55 |
ICC – intraclass correlation coefficients, CI – confidence interval; SEM – standard error of measurement; MDC – minimal detectable change
In this study, we analyzed the inter-machine reliability for both the neutral position and the 60° abduction position. In both positions, the results were excellent in terms of inter-machine reliability. In the US-based evaluation of SIS, the passage of the tendon and subacromial bursa under the acromion is observed. Compression is suspected if fluid in the subacromial bursa collects laterally, the large tubercle does not pass under the acromion, and pain is present during dynamic evaluation(22). However, these findings are usually not observed on the US in individuals clinically diagnosed with SIS(19). Therefore, AHD measurements were performed to evaluate the subacromial space in patients with SIS. In US-based studies reported in the literature, patients with SIS and healthy control groups were compared in terms of AHD, and differences were found(6,7,17). Desmeules
A study on device consistency found excellent inter-machine reliability for patellar tendon length and cross-sectional area(29). Another study found that patellar, Achilles, and plantar fascia thickness could be measured reliably with different US devices(30). However, only SUS devices were used in these studies. An ultrasound-based inter-machine reliability study has recently been published. To measure the cross-sectional area of the median nerve and the thickness of the tendons, SUS and WUS devices were used. Correlation analysis, not intraclass correlation coefficients, was used here to assess the measurement relationship. The measurements had a moderate correlation between 0.43–0.77(31). This study evaluated the application of WUS devices in the musculoskeletal system for the first time, bringing a new perspective to the literature. Future studies may investigate the reliability of WUS in different musculoskeletal regions.
The limitations of our study are the following: 1) intra- and interrater reliability of supraspinatus tendon thickness with the WUS device was not evaluated, and 2) inter-rater reliability of image interpretation was not evaluated.
The findings of our study show that WUS devices are reliable in comparison to SUS devices in the measurement of the acromiohumeral distance in patients with shoulder impingement syndrome. The results of the study could contribute to a significant improvement in the clinical use of WUS devices.