An uncomplicated, sensitive liquid chromatography linked to mass spectrometry (LC/MS) for evaluation of carbamazepine and carbamazepine-10,11-epoxide (its metabolite) in human plasma, human saliva, rat plasma, and rabbit plasma was developed. Analyses were conducted on a Zorbax SB-C18, 100 mm × 3 mm ID, 3.5 μm column, at a column temperature of 40 ºC. The mobile phase was comprised of 0.1% formic acid in water and methanol in a 35 : 65 (v/v) ratio, with a flow rate of 0.4 mL/min. Lacosamide was utilized as internal standard. Under these chromatographic conditions, the retention times of lacosamide, carbamazepine-10,11-epoxide, and carbamazepine were 1.4 min, 1.6 min, and 2.2 min, respectively. The quantification of the analytes was performed using multiple reaction monitoring, with the use of a triple quadrupole mass spectrometer with electrospray positive ionization. The monitored ions were