Comparative study of the reliability of ultrasound to confirm the position of endotracheal tube with cuff inflated with saline versus air


Cite

Fig. 1.

Protocol of study
Protocol of study

Fig. 2.

Comparison of diagnostic certainty (a) and visibility (b) between the 1st and 2ndseries of measurements
Comparison of diagnostic certainty (a) and visibility (b) between the 1st and 2ndseries of measurements

Fig. 3.

LC-CUSUM learning curve for the TA technique
LC-CUSUM learning curve for the TA technique

Sensitivity, specificity, VPP, VPN

Reference Test Sensitivity Specificity VPP VPN
Empirical esophagus 38.5 66.7 89.6 12.6
Empirical trachea 62.2 100 100 26.1

Techniques

Failure N = 154 (53.5%) Success N = 134 (46.5%) p **
N %/mean SD* N %/mean SD*
Endotracheal tube position 0.006
trachea 67 43.5 80 59.7
esophagus 87 56.5 54 40.3
Cuff 1.000
air 77 50.0 67 50.0
saline 77 50.0 67 50.0
Time 154 21.6 7.4 134 19.5 7.8 0.022

Comparison of 1st and 2nd series of measurements

1st series N = 96 (50.0%) 2nd series N = 96 (50.0%) p **
N %/mean SD* N %/mean SD*
Primary outcome 0.004
failure 45 46.9 26 27.1
success 51 53.1 70 72.9
Time 96 21.4 8.0 96 20.6 8.6 0.504

First series of measurements

N %
Endotracheal tube position
Trachea 147 51
Esophagus 141 49
Cuff
Air 144 50
Saline 144 50
Primary outcome
Failure 154 53.5
Success 134 46.5
Diagnostic certainty
I don’t know 34 11.8
Not sure 47 16.3
Pretty sure 72 25.0
Almost sure 82 28.5
Quite sure 53 18.4
Visibility
I see nothing 39 13.5
Poor visibility 45 15.6
Average visibility 73 25.3
Good visibility 95 33.0
Very good visibility 36 12.5