Cite

Aim of the study: At present, there are few scatter particles used in preparing blood-mimicking fluids, such as nylon, sephadex, polystyrene microsphere, and poly(4-methystyrene). In this study, we present cholesterol as a new scatter particle for blood-mimicking fluid preparation. Materials and methods: The procedure for the preparation of the proposed blood-mimicking fluid involved the use of propylene glycol, D(+)-Glucose and distilled water to form a ternary mixture fluid, with cholesterol used as scatter particles. Polyethylene glycol was first used as part of the mixture fluid but the acoustic and physical properties were not suitable, leading to its replacement with D(+)-Glucose, which is soluble in water and has a higher density. A common carotid artery wall-less phantom was also produced to assess the flow properties. Results: The prepared blood-mimicking fluid with new scatter particles has a density of 1.067 g/cm3, viscosity of 4.1 mPa.s, speed of sound 1600 m/s, and attenuation of 0.192 dB/cm at 5 MHz frequency. Peak systolic velocity, end diastolic velocity and mean velocity measurements were gotten to be 40.2 ± 2.4 cm/s, 9.9 ± 1.4 cm/s, and 24.0 ± 1.8 cm/s, respectively. Conclusion: Based on the results obtained, the blood-mimicking fluid was found suitable for ultrasound applications in carotid artery wall-less phantoms because of its good acoustic and physical properties.

eISSN:
2451-070X
Language:
English
Publication timeframe:
4 times per year
Journal Subjects:
Medicine, Basic Medical Science, other