Validity of measurements of testicular volume obtained by a built-in software of ultrasound systems: with formula recommended by updated guidelines as reference


Cite

Fig. 1.

A 31 year-old man with infertility. Measurement of the right testicular volume using ultrasound. The length, width, and height of the testis are 3.91 cm, 2.62 cm, and 2.38 cm, respectively; and the testicular volume is 12.8 mL, which is obtained automatically by the Philips EPIQ5 ultrasound system with a built-in software
A 31 year-old man with infertility. Measurement of the right testicular volume using ultrasound. The length, width, and height of the testis are 3.91 cm, 2.62 cm, and 2.38 cm, respectively; and the testicular volume is 12.8 mL, which is obtained automatically by the Philips EPIQ5 ultrasound system with a built-in software

Comparison of values obtained by the built-in software of ultrasound systems and re-calculated values with reference to the updated guidelines

Ultrasound systems Manufacture date Software update date Distribution of testicular measurement (n = 1909) Concordance after validity (n = 566) Discordance after validity (n = 1343)
Siemens Acuson S2000 September 2009 July 2015 352 0 352
Philips EPIQ5 February 2018 No 224 0 224
GE Logiq E9 March 2017 No 362 0 362
Mindray DC-8 June 2012 January 2017 363 363 0
Mindray Resona 7 November 2017 No 203 203 0
Hitachi Aloka Prosound α7 October 2010 No 505 0 505