Implementation of advanced radiation therapy techniques into clinical practice has set high demands on the quality and accuracy of various devices used for radiation treatment planning, treatment delivery, and dose verification. Besides the required high performance of medical linear accelerators and their ancillary systems, there are also strict requirements on dose calculation and optimization using treatment planning systems (TPS). Precise dose calculation is one of the most critical steps in the radiation therapy process since it is the basis for accurate and safe treatment delivery using high-energy photon beams. To provide necessary dosimetric accuracy, the verification of the calculated doses should be performed using a reproducible and reliable methodology. To ensure acceptable reliability of the verification results, an appropriate methodology for dose verification should be carefully selected, while the limitations of the specific method must be fully understood. The latter is essential for an adequate interpretation of the verification results.
Comprehensive verification methodology for the evaluation of calculation algorithms built in the TPSs has been proposed by the International Atomic Energy Agency (IAEA).1, 2 However, the rapid development of treatment delivery devices and, consequently, the utilization of the advanced radiation therapy techniques call for further development of the verification methods. In some published studies and documents3, 4, 5, methodologies for the verification of dosimetry parameters for the implementation of Intensity Modulated Radiotherapy (IMRT) have been proposed. However, neither the means of verification nor the methods were explicitly spelled out.
Presently, Monte Carlo based dose calculation algorithms built in the TPS are assumed to be the most accurate computational systems for the appropriate simulation of particle transport and dose calculation.6,7 Those algorithms offer two alternative options for the calculation and reporting of the absorbed dose: dose-to-medium as calculated by Monte Carlo algorithms, referred as dose to medium-in-medium,
Since it is a matter of debate whether to use
The ultimate goal of the study was to define and propose an additional verification procedure as a supplement to the set of existing preclinical commissioning tests provided in the IAEA TECDOC 15832, for the specific case where TPS uses Monte Carlo based calculation algorithms. Such additional test may well eliminate potential misinterpretations of the commissioning results for bone density material, where
We have to note that the proposed addendum to the verification methodology has no intention to be an answer to which reporting mode,
In this work we used 6 MV photon beam generated by Siemens Oncor Expression (Siemens Healthineers, Erlangen, Germany) linear accelerator, Siemens Somatom Open Computerized Tomography (CT) simulator (Siemens Helthineers, Erlangen, Germany) and Elekta Monaco treatment planning system version 5.11 (Elekta, Stockholm, Sweden). Monaco TPS is a Monte Carlo based system which calculates absorbed dose using the
Accuracy of the TPS Monaco ver. 5.11 calculation algorithm was experimentally verified using a semi-anthropomorphic CIRS Thorax phantom (CIRS Inc., Norfolk, VA, USA) consisting of a body made of water equivalent material (ρ = 1.003 g/cm3), lung equivalent parts (ρ = 0.207 g/cm3), and bone equivalent part (ρ = 1.506 g/cm3) with cylindrical holes for placement of ionization chamber into interchangeable rod inserts having three different densities.2 The phantom was scanned using the Somatom Open CT simulator. Acquired CT images were used for the delineation of volumes of interest and subsequent dose calculations. Measurements of absorbed dose were performed at ten measuring positions within the phantom (Figure 1) for 15 different irradiation set-ups (Table 1), using a PTW Farmer-type ionization chamber. All measurements were carried out at the central part of the selected radiation fields, excluding the regions of high dose gradients.
Figure 1
Photo of the semi-anthropomorphic CIRS Thorax phantom with interchangeable rod inserts (left) and its CT image (right). Positions of 10 interchangeable rod inserts are marked with numbers from 1 to 10. Five measuring points are in the water equivalent part of the phantom (grey area), four points are in the lung density equivalent material (black area), and one point is in the bone density equivalent part of the phantom (white area).

Irradiation set-ups for measurements in 6 MV photon beam used for experimental verification of the Monaco ver. 5.11 treatment planning systems (TPS) calculation algorithm in the semi-anthropomorphic CIRS Thorax phantom. Reference and measuring points (I1 to I10) are shown in the last two columns; subscripts 1 to 10 correspond to the labelling in Figure 1
Set-up | Irradiation geometry | Field size [cm2] | SSD/SAD | Gantry angle [°] | reference point | measuring points |
---|---|---|---|---|---|---|
1 | Single square fields | 10×10 | SSD | 0 | I5 | I1, I3, I5-10 |
2 | 10×10 | SAD | 0 | I5 | I, I, I135-10 | |
3 | 4×4 | SAD | 0 | I5 | I1-9 | |
4 | 10×10 | SAD | 90 | I3 | I2-10 | |
5 | Rectangular field | 10× 15 | SAD | 300 | I1 | I1, I4, I6-8, I10 |
6 | Single asymmetric fields | (6+8)×15 | SAD | 0 | I5 | I1-10 |
7 | (3+8)×15 | SAD | 90 | I5 | I, I15-10 | |
8 | (4+10)×15 | SAD | 180 | I5 | I1-3, I5-10 | |
9 | (3+7)×15 | SAD | 300 | I5 | I2-10 | |
12×10 | SAD | 0 | ||||
10 | 4 fields (box) | 12×10 | SAD | 180 | I5 | I2-5 |
12×8 | SAD | 90 | ||||
12×8 | SAD | 270 | ||||
4×4 | SAD | 30 | ||||
11 | 3 fields | 16×4 | SAD | 90 | I5 | I2, I5-9 |
16×4 | SAD | 270 | ||||
12 | Diamond-shaped field | 14×14 | SAD | 0 | I3 | I1, I3, I5-10 |
13 | Irregular L shaped field | / | SAD | 45 | I1 | I1-2, I4-6, I8-10 |
14 | MLC cylinder shaped field | / | SAD | 0 | I2 | I1,2, I5, I8,9, I10 |
16×4 | SAD | 90 | ||||
15 | 3 non-coplanar fields | 16×4 | SAD | 270 | I5 | I1, I5-6, I8, I10 |
4×4 Couch angle = 270° | SAD | 30 |
SAD = source to axis distance; SSD = source to surface distance
Measured doses were compared to the corresponding doses obtained by both calculation options,
where
Dose differences
The index
Throughout the study, all calculations within Monaco TPS were performed on a 0.2 cm calculation grid, with 0.5% statistical uncertainty per control point.
In the second part of the study, we were aiming to determine differences between
Three irradiation geometries (single asymmetric rectangular fields having different gantry angles: 0°, 90°, and 180°) were selected for this part of the study (Table 1, set-ups 6, 7, and 8). For each of those irradiation geometries, two phantom assemblies were used to analyze differences between the two calculation approaches with respect to the measurements performed by PTW 30013 Farmer type ionization chamber in the bone density equivalent (BDE) part of the CIRS Thorax phantom. In the first assembly, referred to as
Figure 2
CT image of the CIRS Thorax phantom: water equivalent insert inside BDE part of the phantom

In the last part of the study, the phantom assembly was additionally virtually modified for the calculation purposes in the Monaco TPS: cylinders of various volumes (constant length and different diameters) were delineated inside the BDE insert on the CT scans (Figure 2, top right). This approach was utilized to obtain the limits above which the differences between
To verify the accuracy of dose-to-medium and dose-to-water calculation modes, we have analyzed differences
The uncertainty of
We considered that the
was satisfied.
Secondly, we considered that the dose calculations within Monaco TPS were in agreement with the experimentally determined doses if the conditions
were satisfied. At this point we note, that throughout the rest of the paper all combined uncertainties are stated within two standard deviations,
Differences between calculated and measured absorbed doses for two calculation modes, dose-to-medium
Figure 3
Mean percentage dose differences

Comparison of measured and calculated doses in the water equivalent part of the phantom showed that the mean percentage dose difference for all points was - 0.6%
Comparison of measured and calculated doses in the lung density equivalent part of the phantom showed that
In the bone density equivalent part of the CIRS Thorax phantom, the percentage dose differences between the two calculation options were larger than in the previous two cases (Figure 3). Mean difference
Dose calculations within Monaco TPS were in agreement with experimentally determined doses for water equivalent and lung equivalent parts of the CIRS Thorax phantom, since the conditions from Eqs. [6] and [7] were satisfied. On the contrary, for the BDE part of the phantom, conditions from Eqs. [6] and [7] were not satisfied. Therefore, we can conclude that the dose calculations in Monaco TPS ver. 5.11 were not in agreement with measured absorbed doses for the BDE part of the phantom, regardless of the calculation mode.
The second part of the study was focused on the differences between calculated and measured absorbed doses in the BDE part of the CIRS Thorax phantom. Three simple asymmetric fields with different gantry angles were selected for that purpose utilizing two different phantom assemblies,
Differences
Irradiation geometry (field, gantry) | Phantom assembly | ||
---|---|---|---|
(6+8) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 2.9 | 2.9 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | - 0.7 | - 0.2 | |
(3+8) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 3.0 | 5.1 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | - 0.7 | - 0.1 | |
(4+10) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 5.7 | 5.4 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | 0.5 | 1.3 |
For
In the
The absolute value of the difference between both approaches was in this case statistically significant:
As a final point, we investigated differences between calculated and measured doses in the phantom, which was virtually modified for the calculation purposes, as described in the section Materials and methods. Results for five delineated “water cylinders,” including the results for
Mean differences,
V [cm3] | ||||
---|---|---|---|---|
0 | - 3.9 | 2.1 | 4.4 | 1.9 |
0.035 | - 2.6 | 1.5 | 2.5 | 1.9 |
0.141 | - 1.4 | 1.3 | 1.8 | 1.7 |
0.279 | - 0.3 | 1.2 | 1.2 | 1.5 |
0.573 | 0.3 | 1.4 | 0.4 | 1.3 |
Differences between calculated and measured doses in the water equivalent part of the CIRS Thorax semi-anthropomorphic phantom were within 1% and not significantly different from zero (Eqs. [6] and [7]), regardless of the applied calculation option. The latter is in good agreement with previously published data.9,16 Similarly, in lung density equivalent material, the calculated mean percentage dose differences were not significantly different than zero for both calculation modes, confirming the results from previously published studies.3,9,13
The differences between the two calculation approaches, dose-to-medium and dose-to-water, were, however, significant in BDE media (Table 2 and Figure 3). Andreo
In the BDE part of the CIRS Thorax semi-anthropomorphic phantom, mean percentage dose differences
However, differences between the respective mean values
For further discussion, analysis, and graphical presentation, the exponential function was selected to fit the data from Table 3. The general form of the fitting function is given as
with fitting coefficients,
for
Figure 4
Average differences

Applying Eqs. [9] and[10] for large volumes, we can see that
Differences
Irrespective of the fact that the ionization chamber is calibrated in terms of dose to water, we propose an additional verification test of the accuracy of the Monaco TPS calculation modes for BDE regions considering the mentioned observations:
One can select three simple irradiation geometries (single fields, different gantry angles) and perform measurements of absorbed doses with the Farmer type ionization chamber in the BDE part of CIRS Thorax semi-anthropomorphic phantom, using a BDE insert (“standard” geometry). The ionization chamber should be positioned at the central part of the radiation field, where the measured signal is sufficiently large.
Measured doses are compared to the calculated ones using both calculation modes,
Obtained mean values
Finally, the acceptability of the tested TPS algorithm is based on two conditions, which have to be fulfilled concurrently:
Differences
Fulfilment of this condition means that both calculation options yield to the same results within statistical uncertainty for large volumes, as expected. Since significant differences do exist for small volumes of delineated “water cylinders,” we have to consider this fact as well. The maximal difference
Obtained values
It is important to note that our investigation was limited to the region of charged particle equilibrium (CPE) and for 6 MV photon beam only.
In the present study, a Monte Carlo based calculation algorithm built in the Elekta Monaco ver. 5.11 TPS was analyzed for 6 MV photon beam. It was confirmed that both calculation approaches, dose-to-medium and dose-to-water, yield to the similar results in the water equivalent and lung density equivalent parts of the semi-anthropomorphic phantom and are in agreement with experimentally determined absorbed doses.
In the bone density equivalent part of the phantom, significant differences were observed when calculations were compared to the measured absorbed doses. While the dose-to-medium approach yields to lower doses compared to the measured ones, calculations utilizing the dose-to-water computing approach revealed similar differences but of opposite sign. The observed differences can lead to ambiguity regarding the acceptability of the verification results before the clinical implementation of a newly commissioned TPS Monaco.
To overcome the ambiguity on the pertinence of the verification results in the bone density equivalent material, a supplement to the current TPS commissioning methodology has been proposed, having in mind inherent differences between the two calculation modes. This supplement relies on the findings from the present study. We consider it as a consistent and efficient method for the experimental verification of the absorbed dose calculation in both calculation modes
While the selection of beams in this study assumes conditions of charged particle equilibrium, it would be highly interesting and worthwhile to set-up the study where CPE is violated, e.g., for small fields where lateral CPE does not exist. However, an experimental determination of absorbed doses in small fields is demanding. It requires determination of detector specific correction factors, which have to be utilized individually for the selected detector and are associated with additional uncertainties.24, 25, 26 The latter can pose a problem to conduct such a study with sufficient reliability and robustness.
Figure 1

Figure 2

Figure 3

Figure 4
![Average differences δD¯m${{\overline{\delta D}}_{m}}$and δD¯w${{\overline{\delta D}}_{w}}$between calculated and measured doses in the bone density equivalent (BDE) part of the CIRS Thorax phantom, as a function of the volumes of the simulated “water cylinders” (see Figure 2 and Table 3). δD¯m${{\overline{\delta D}}_{m}}$and δD¯w${{\overline{\delta D}}_{w}}$are presented as individual values/points calculated using Eqs. [1] to[4], and in the form of two analytical functions from Eqs. [9] and [10]. Error bars represent corresponding uncertainties within 95% confidence limits.](https://sciendo-parsed.s3.eu-central-1.amazonaws.com/647355d44e662f30ba53aa4b/j_raon-2020-0051_fig_004.jpg?X-Amz-Algorithm=AWS4-HMAC-SHA256&X-Amz-Date=20231201T041702Z&X-Amz-SignedHeaders=host&X-Amz-Expires=18000&X-Amz-Credential=AKIA6AP2G7AKP25APDM2%2F20231201%2Feu-central-1%2Fs3%2Faws4_request&X-Amz-Signature=7b11b7a030aaef10e038455ce59cf6fbb912a19af756e9f67d8095e91a8100f0)
Differences δDm$\delta {{D}.{m}}$and δDw$\delta {{D}.{w}}$between two different calculation options in the Monaco ver. 5.11. treatment planning systems (TPS) and measured data obtained in the bone density equivalent (BDE) part of the CIRS Thorax phantom, according to Eqs. [1] and [2]. Two phantom assemblies and three simple beam setups were considered for this part of the study
Irradiation geometry (field, gantry) | Phantom assembly | ||
---|---|---|---|
(6+8) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 2.9 | 2.9 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | - 0.7 | - 0.2 | |
(3+8) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 3.0 | 5.1 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | - 0.7 | - 0.1 | |
(4+10) x 15 cm2 | standard BDE insert with the ionization chamber placed in the BDE part of the phantom | - 5.7 | 5.4 |
non-standard Water equivalent insert with the ionization chamber placed in the BDE part of the phantom | 0.5 | 1.3 |
Mean differences, δD¯m${{\overline{\delta D}}.{m}}$and δD¯w${{\overline{\delta D}}.{w}}$between calculated and measured doses in the bone density equivalent (BDE) part of the CIRS Thorax phantom for Dm and Dw calculation approaches, respectively. The absorbed doses were calculated using the Monaco ver. 5.11 treatment planning systems (TPS) in the center of delineated “water cylinders” of volume V, in the BDE part of the phantom. Corresponding combined uncertainties are denoted as uc,m and uc,w for dose-to-medium and dose-to-water calculation options, respectively
V [cm3] | ||||
---|---|---|---|---|
0 | - 3.9 | 2.1 | 4.4 | 1.9 |
0.035 | - 2.6 | 1.5 | 2.5 | 1.9 |
0.141 | - 1.4 | 1.3 | 1.8 | 1.7 |
0.279 | - 0.3 | 1.2 | 1.2 | 1.5 |
0.573 | 0.3 | 1.4 | 0.4 | 1.3 |
Irradiation set-ups for measurements in 6 MV photon beam used for experimental verification of the Monaco ver. 5.11 treatment planning systems (TPS) calculation algorithm in the semi-anthropomorphic CIRS Thorax phantom. Reference and measuring points (I1 to I10) are shown in the last two columns; subscripts 1 to 10 correspond to the labelling in Figure 1
Set-up | Irradiation geometry | Field size [cm2] | SSD/SAD | Gantry angle [°] | reference point | measuring points |
---|---|---|---|---|---|---|
1 | Single square fields | 10×10 | SSD | 0 | I5 | I1, I3, I5-10 |
2 | 10×10 | SAD | 0 | I5 | I, I, I135-10 | |
3 | 4×4 | SAD | 0 | I5 | I1-9 | |
4 | 10×10 | SAD | 90 | I3 | I2-10 | |
5 | Rectangular field | 10× 15 | SAD | 300 | I1 | I1, I4, I6-8, I10 |
6 | Single asymmetric fields | (6+8)×15 | SAD | 0 | I5 | I1-10 |
7 | (3+8)×15 | SAD | 90 | I5 | I, I15-10 | |
8 | (4+10)×15 | SAD | 180 | I5 | I1-3, I5-10 | |
9 | (3+7)×15 | SAD | 300 | I5 | I2-10 | |
12×10 | SAD | 0 | ||||
10 | 4 fields (box) | 12×10 | SAD | 180 | I5 | I2-5 |
12×8 | SAD | 90 | ||||
12×8 | SAD | 270 | ||||
4×4 | SAD | 30 | ||||
11 | 3 fields | 16×4 | SAD | 90 | I5 | I2, I5-9 |
16×4 | SAD | 270 | ||||
12 | Diamond-shaped field | 14×14 | SAD | 0 | I3 | I1, I3, I5-10 |
13 | Irregular L shaped field | / | SAD | 45 | I1 | I1-2, I4-6, I8-10 |
14 | MLC cylinder shaped field | / | SAD | 0 | I2 | I1,2, I5, I8,9, I10 |
16×4 | SAD | 90 | ||||
15 | 3 non-coplanar fields | 16×4 | SAD | 270 | I5 | I1, I5-6, I8, I10 |
4×4 Couch angle = 270° | SAD | 30 |